Accuracy of automated and non-audiologist-operated audiometry compared to gold-standard testing
Carl Pedersen1,2,3, Ellen Raben Pedersen4, Chris Bang Sørensen4
1Research Unit for ORL - Head & Neck Surgery and Audiology, Odense University Hospital, Odense, Denmark.
Objective:
This study compared the accuracy of automated user-operated and manual non-audiologist driven audiometry, two potential alternatives for test settings with limited audiologist resources. The accuracy was investigated by assessing the two test methods with gold-standard manual audiometry conducted by trained audiologists.
Design:
This comparative study was conducted in a clinical setting. The audiometry conducted by non-audiologist operators took place at the participants' local Ear, Nose, and Throat specialists. Both the user-operated and gold-standard audiometry were conducted on the same day at Odense University Hospital, approximately one month after the initial assessment using the non-audiologist driven audiometry.
Study Sample:
221 adults (median age of 69 years) referred for hearing-aid treatment.
Results:
Both test methods showed statistically significant differences from the gold standard, but the user-operated audiometry demonstrated closer agreement at all tested frequencies. Across frequencies, user-operated audiometry differed by -0.1 dB (95% CI -0.9 to 0.8) to 2.6 dB (95% CI 1.7-3.5), while non-audiologist driven audiometry differed by 6.2 dB (95% CI 5.5-6.9) to 11.1 dB (95% CI 10.0-12.3).
Conclusion:
The accuracy of automated user-operated audiometry was within general measurement uncertainty for gold-standard audiometry, whereas the manual non-audiologist driven audiometry may introduce additional variability.
More Related Videos
03:58Enhancing Electrode Location Assessment in Cochlear Implantation via Computed Tomography Image Fusion
Published on: January 17, 2025
14:05Behavioral Assessment of Hearing in 2 to 4 Year-old Children: A Two-interval, Observer-based Procedure Using Conditioned Play-based Responses
Published on: January 23, 2017
Related Concept Videos
Testing a Claim about Standard Deviation
The hypothesis testing for the claim of population standard deviation (or variance) requires the data and samples to be random and unbiased. The population distribution also must be normal. There is no specific requirement on the sample size as the estimation is based on the chi-square distribution.
As a first step, the hypothesis (null and alternative) concerning the claim about...
Accuracy and Errors in Hypothesis Testing
In hypothesis testing, the probability of making a Type I error, denoted as α, is commonly set at 0.05. This significance level indicates a 5%...