Laboratory Evaluation of a Wearable Instrumented Headband for Rotational Head Kinematics Measurement
Anu Tripathi1, Yang Wan2, Sushant Malave3
1Department of Engineering, Robert Morris University, Moon Township, PA, USA.
Purpose:
Mild traumatic brain injuries (mTBI) are a highly prevalent condition with heterogeneous outcomes between individuals. A key factor governing brain tissue deformation and the risk of mTBI is the rotational kinematics of the head. Instrumented mouthguards are a widely accepted method for measuring rotational head motions, owing to their robust sensor-skull coupling. However, wearing mouthguards is not feasible in all situations, especially for long-term data collection. Therefore, alternative wearable devices are needed. In this study, we present an improved design and data processing scheme for an instrumented headband.
Methods:
Our instrumented headband utilizes an array of inertial measurement units (IMUs) and a new data processing scheme based on continuous wavelet transforms to address sources of error in the IMU measurements. The headband performance was evaluated in the laboratory on an anthropomorphic test device, which was impacted with a soccer ball to replicate soccer heading.
Results:
When comparing the measured peak rotational velocities (PRV) and peak rotational accelerations (PRA) between the reference sensors and the headband for impacts to the front of the head, the correlation coefficients (r) were 0.80 and 0.63, and the normalized root mean square error (NRMSE) values were 0.20 and 0.28, respectively. However, when considering all impact locations, r dropped to 0.42 and 0.34 and NRMSE increased to 0.5 and 0.41 for PRV and PRA, respectively.
Conclusion:
This new instrumented headband improves upon previous headband designs in reconstructing the rotational head kinematics resulting from frontal soccer ball impacts, providing a potential alternative to instrumented mouthguards.
More Related Videos
07:30A Test Bed to Examine Helmet Fit and Retention and Biomechanical Measures of Head and Neck Injury in Simulated Impact
Published on: September 21, 2017
06:52An Inertial Measurement Unit Based Method to Estimate Hip and Knee Joint Kinematics in Team Sport Athletes on the Field
Published on: May 26, 2020
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...