Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
Published on: January 18, 2021
Idealized aortic annuloplasty FSI digital twin of 3D-printed phantoms with 4D-flow MRI comparison
Luca Bontempi1, Marta Zattoni2, Anna Ramella3
1Department of Mechanical and Production Engineering, Aarhus University, Aarhus, Denmark; Laboratory of Biological Structure Mechanics (LaBS), Department of Chemistry, Materials and Chemical Engineering "Giulio Natta", Politecnico di Milano, Milan, Italy; Department of Biomedical Engineering, Thoraxcenter, Erasmus MC, Rotterdam, the Netherlands.
Background:
Aortic annuloplasty, involving the implantation of an external ring around the aortic root to reduce annular dimensions, is a promising treatment for aortic valve insufficiency. However, its hemodynamic effects remain underexplored due to the absence of computational models validated by experimental and clinical data.
Methods:
This study introduces a computational fluid-structure interaction (FSI) model of supra valvular aortic annuloplasty using 4D-flow magnetic resonance imaging (MRI). Native and post-annuloplasty conditions of idealized aortic root phantoms, including the aortic valve, were CAD-modelled and 3D-printed with elastic resin. These phantoms were tested in a mock circulatory flow-loop providing normal pulsatile physiologic conditions using a glycerol-water mixture to simulate blood viscosity. Flow and pressure data collected from sensors were used as boundary conditions for FSI simulations. Experimental velocity fields from 4D-flow MRI were compared to computational results to assess model accuracy.
Results:
MRI scans of the annuloplasty model showed an increased peak systolic velocity (up to 145.4 cm/s) and localized flow alterations, corresponding to a higher pressure gradient across the valve. During regurgitation, the annuloplasty model showed broader velocity distributions compared to the native condition. The FSI simulations closely matched 4D-flow MRI data, with strong correlation coefficients (r > 0.93) and minimal Bland-Altman differences, particularly during systolic phases.
Conclusions:
This study establishes an integrative methodology combining in-vitro, in-silico, and clinical imaging techniques to evaluate aortic annuloplasty hemodynamics. The in-vitro validated digital twin framework offers a pathway for patient-specific modelling, enabling prediction of surgical outcomes and optimization of aortic valve repair strategies.