Related Experiment Video
Updated: May 21, 2025

Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
Re-calculation of air kerma to dose-equivalent conversion coefficients for mono-energetic photons
Jaroslav Šolc1, Paz Aviles Lucas2, Luka Bakrač3
1Czech Metrology Institute (CMI), Okruzni 31, 638 00 Brno, Czech Republic.
Abstract:
This work presents new calculations of conversion coefficients (CCs) from total air kerma to dose-equivalent quantitiesH'(0.07),H'(3),H*(10),Hp(0.07),Hp(3), andHp(10) for area and personal dosimetry for mono-energetic photons in the energy range from 2 keV to 50 MeV, assuming secondary charged particle equilibrium. Calculations using the Monte Carlo N-Particle® (MCNP) code were performed for a large number of photon energies with the aim of preventing errors resulting from possible improper interpolation between currently available sparsely spaced CC values, when an average CC value over a photon fluence spectrum needs to be determined. The CC values were compared with the values published in the ISO 4037-3:2019 standard. A very close agreement was achieved for the majority of CCs. Larger discrepancies were found for some CCs, often for low photon energies or large angles of radiation incidence, which were taken from older publications or when CC values were interpolated or extrapolated. Furthermore, some differences were found in the MeV energy range, which are significant for dosimeter calibrations, e.g. the presented values of CC toH*(10) for the main photon energies of137Cs and60Co radionuclides are both lower by 2.8%. Finally, it was found that the values of CCs toHp(0.07;E, α)slabgiven in ISO 4037-3:2019 were not taken correctly from the source publication. In conclusion, the CC values given in ISO 4037-3:2019 should be updated in view of the results obtained.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Related Concept Videos
Biological Effects of Radiation
Absorption of Radiation
One-Compartment Open Model: Wagner-Nelson and Loo Riegelman Method for ka Estimation
On...
Conduction, Convection and Radiation: Problem Solving
In order to solve a problem related to heat transfer, first of all, the situation needs to be examined to determine the type of heat transfer involved. This could...
Radiation Pressure: Problem Solving
The average value of the rate of momentum transfer divided by the absorbing area represents the average force...
Homogeneous Equilibria for Gaseous Reactions
For gas-phase reactions, the equilibrium constant may be expressed in terms of either the molar concentrations (Kc) or partial pressures (Kp) of the reactants and products. A relation between these two K values may be simply derived from the ideal gas equation and the definition of molarity. According to the ideal gas equation: