Implementation of the ISO 4037:2019 standard using voltage dividers and X-ray spectrometry

Luigi Rinaldi1, Alessia Ciccotelli1, Claudia Silvestri1

  • 1Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti, ENEA-INMRI, Centro Ricerche ENEA Casaccia, Via Anguillarese 301, 00123 Santa Maria di Galeria, Roma, Italy.

Abstract

Related Concept Videos

Voltage Dividers01:14

Voltage Dividers

In electrical circuits, resistors can be connected in series, sequentially linked one after the other. In a series configuration, the same current flows through each resistor. Ohm's law is a fundamental principle to understand the behavior of resistors in series. It expresses the voltage across these resistors in terms of the current and resistance.
Kirchhoff's voltage law implies that the sum of the voltages across the resistors in series equals the source voltage. This means that the current...
IR Spectrometers01:25

IR Spectrometers

There are two main infrared (IR) spectrophotometers: dispersive IR spectrometers and Fourier transform infrared (FTIR) spectrometers. In a dispersive IR spectrometer, a beam of infrared radiation produced by a hot wire is divided into two parallel equal-intensity beams using mirrors. One beam passes through the sample, while another is a reference beam. The beams then move through the monochromator, which separates the radiations into a continuous spectrum of different frequencies. The...
UV–Vis Spectrometers01:14

UV–Vis Spectrometers

The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell. Samples for...
Atomic Absorption Spectroscopy: Instrumentation01:22

Atomic Absorption Spectroscopy: Instrumentation

An atomic absorption spectrophotometer (AAS) comprises several components: a radiation source, an atomizer, a monochromator, and a detector. The radiation source can be a hollow-cathode lamp (HCL) or an electrodeless-discharge lamp (EDL), both of which provide a narrow emission line of the required wavelength. However, some instruments use continuum sources and high-resolution monochromators to achieve a narrow range of radiation.
The atomizer used in AAS can be either a flame atomizer or an...
Voltammetry: Factors Affecting Measurements01:21

Voltammetry: Factors Affecting Measurements

A current produced due to the redox reactions of the analyte at the working and auxiliary electrodes is called a faradaic current. The reaction can be divided into two types. The current generated due to the reduction of the analyte is called cathodic current, and it carries a positive charge. In contrast, the current produced by analyte oxidation is known as an anodic current, and it has a negative charge. The applied potential at the working electrode determines the faradaic current flow, and...
Atomic Absorption Spectroscopy: Lab01:21

Atomic Absorption Spectroscopy: Lab

For AAS measurements, samples must be introduced as clear solutions, often requiring extensive preliminary treatment to dissolve materials like soils, animal tissues, and minerals. Common methods for sample preparation include treatment with hot mineral acids, wet ashing, combustion in closed containers, high-temperature ashing, or fusion with reagents.
 Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...