Position-Dependent and Hidden Atrioventricular Dyssynchrony in Micra AV Leadless Pacemaker
1Department of Cardiology, Soka Municipal Hospital, Soka, Japan.
Pacing and Clinical Electrophysiology : PACE
|May 20, 2025
Summary
Leadless pacemakers like Micra AV can cause position-dependent issues. Careful analysis of signal variations and histograms is crucial for diagnosing and resolving symptoms.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Device Technology
Background:
- The Micra AV leadless pacemaker is designed to maintain atrioventricular (AV) synchrony in patients.
- AV synchrony is critical for effective cardiac function and patient well-being.
Observation:
- An 81-year-old male patient experienced nocturnal palpitations.
- Routine evaluations were normal, but rate histogram analysis showed an unusual pattern.
- Manual atrial mechanical (MAM) testing revealed position-dependent signal oversensing.
Findings:
- The patient exhibited position-dependent AV dyssynchrony, specifically oversensing in the left lateral position.
- This oversensing was linked to an increased A3 signal amplitude.
- Adjusting the A3 threshold and window end settings successfully resolved the patient's symptoms.
Implications:
- This case underscores the significance of evaluating body position-related signal variations in leadless pacemaker function.
- Rate histogram analysis is a valuable tool for identifying subtle pacemaker behavior issues.
- Optimizing pacemaker settings based on position-dependent findings can improve patient outcomes and device efficacy.
Keywords:
Micra AVatrioventricular dyssynchronymanual atrial mechanical testposition‐dependentrate histogramMore Related Videos
Related Concept Videos
Conduction System of the Heart
4.0K
Autorhythmicity is a term that refers to the heart's inherent ability to generate electrical signals and instigate muscle contractions. This self-regulating conduction system within the heart consists of two key components: the pacemaker cells and specialized conducting cells.
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
4.0K
Mechanism of Cardiac Arrhythmias
864
Arrhythmias are irregular heart rhythms occurring when the heart's electrical impulses become abnormal. These disturbances can lead to various symptoms, depending on their severity and the underlying cause. Some common factors contributing to arrhythmias include hypoxia, ischemia, electrolyte imbalances, excessive catecholamine exposure, drug toxicity, and muscle overstretching. Arrhythmias can be classified into two main types based on the rate and site of origin of abnormal heart rhythms.
864
Disturbances in Heart Rhythm
785
Arrhythmia or dysrhythmia refers to an abnormal heart rhythm caused by a defect in the heart's conduction system. It can cause the heart to beat irregularly, too quickly, or too slowly, leading to symptoms like chest pain, shortness of breath, and fainting. Factors such as stress, caffeine, alcohol, nicotine, cocaine, certain drugs, congenital defects, diseases, and electrolyte abnormalities can trigger arrhythmias.
Arrhythmias are categorized by their speed, rhythm, and origin. A slow...
Arrhythmias are categorized by their speed, rhythm, and origin. A slow...
785
Electrophysiology of Normal Cardiac Rhythm
1.9K
The normal cardiac rhythm is a synchronized electrical activity that facilitates the regular and coordinated contraction of the heart muscle. This process is essential for efficient blood circulation throughout the body. The fundamental elements involved in establishing and maintaining this rhythm include the unique electrical properties of cardiac muscle cells, the sinoatrial (SA) node's pacemaker function, the specialized conducting system, and the ionic mechanisms underlying each phase...
1.9K
Antiarrhythmic Drugs: Class IV Agents as Calcium Channel Blockers
746
Class IV antiarrhythmic drugs, such as verapamil and diltiazem, block calcium channels. They primarily affect the heart, slowing the conduction in calcium-dependent tissues like the SA and AV nodes. These drugs manage reentrant supraventricular tachycardia (SVT) and reduce ventricular rate in atrial flutter/fibrillation.
Verapamil, a calcium channel blocker, inhibits calcium movement across myocardial cell membranes and vascular smooth muscle. This results in the dilation of coronary and...
Verapamil, a calcium channel blocker, inhibits calcium movement across myocardial cell membranes and vascular smooth muscle. This results in the dilation of coronary and...
746