Related Experiment Video
Updated: May 23, 2025

08:08
Simulator Training for Endovascular Neurosurgery
Published on: May 6, 2020
3.5K
Macrowire-only direct aspiration first-pass technique for endovascular mechanical thrombectomy: Multicenter technical
William Ares1, Daniel Tonetti2, Manisha Koneru2
1Department of Neurosurgery, Endeavor Health, Evanston, IL, USA.
Summary
The Aristotle Colossus Guidewire safely and effectively facilitates microcatheter-free delivery of large-bore aspiration catheters for mechanical thrombectomy in ischemic stroke patients. This macrowire approach improves procedural efficiency and outcomes in neurovascular interventions.
Area of Science:
- Neurology
- Interventional Neuroradiology
- Medical Devices
Background:
- Large-bore aspiration catheters (ACs) are crucial for mechanical thrombectomy in acute ischemic stroke.
- Microcatheter-free delivery of ACs can streamline endovascular procedures.
- Macrowires may offer a solution for facilitating AC navigation in complex neurovasculature.
Purpose of the Study:
- To evaluate the safety and efficacy of the Aristotle Colossus Guidewire for microcatheter-free delivery of large-bore ACs.
- To assess the performance of this macrowire in anterior circulation endovascular mechanical thrombectomy.
- To identify factors influencing successful AC delivery using the guidewire.
Main Methods:
- Retrospective analysis of prospective data from 12 institutions.
- Inclusion of patients undergoing anterior circulation mechanical thrombectomy with the Aristotle Colossus Guidewire.
- Assessment of primary endpoint: successful AC delivery without adjunct devices; secondary endpoint: successful reperfusion.
Main Results:
- Successful microcatheter-free AC delivery was achieved in 87.7% of 122 patients.
- 76.2% of patients achieved successful reperfusion (TICI grade ≥ 2B) without a microcatheter.
- Severe carotid artery tortuosity and intracranial atherosclerotic disease were associated with lower success rates.
Conclusions:
- The Aristotle Colossus Guidewire demonstrates favorable performance in anterior circulation ischemic stroke intervention.
- This macrowire facilitates safe navigation of large-bore ACs without requiring a microcatheter or stentriever.
- The findings support the use of this macrowire to improve efficiency in mechanical thrombectomy procedures.