Radiation dose in 1-60-month children undergoing 64-detector cardiac CT angiography: ECG-gated versus non-gated scans
T Masuda1, T Nakaura2, Y Funama3
1Department of Radiological Technology, Faculty of Health Science and Technology, Kawasaki University of Medical Welfare, 288, Matsushima, Kurashiki, Okayama, 701-0193, Japan.
Insights
Retrospective electrocardiogram (ECG)-gated cardiac computed tomography angiography (CTA) in pediatric patients resulted in approximately 10 times higher radiation dose compared to non-gated scans. Image quality and noise levels remained comparable between the two methods.
Area of Science:
- Radiology
- Pediatric Imaging
- Radiation Dose Optimization
Background:
- Cardiac computed tomography angiography (CTA) is essential for pediatric cardiovascular assessment.
- Understanding radiation exposure differences between ECG-gated and non-gated helical scans is critical for patient safety.
Purpose of the Study:
- To investigate and compare the radiation exposure dose of ECG-gated versus non-gated helical scans in pediatric cardiac CTA.
- To evaluate the impact of scan type on image quality parameters.
Main Methods:
- A retrospective analysis of 63 pediatric patients (1-60 months) undergoing cardiac CTA.
- Patients were divided into non-gated helical (33) and retrospective ECG-gated (30) scan groups using a 64-detector-row CT.
- Comparison of CT dose-volume index (CTDIvol), dose-length product (DLP), effective dose (ED), CT values, and image noise.
Main Results:
- ECG-gated scans showed approximately 10 times higher radiation doses (CTDIvol, DLP, ED) than non-gated scans (P < 0.01).
- No statistically significant differences were observed in CT values of the aorta and pulmonary artery between the groups (P < 0.05).
- Image noise levels were comparable between retrospective ECG-gated and non-gated helical scans (P < 0.05).
Conclusions:
- Retrospective ECG-gated cardiac CTA significantly increases radiation exposure in pediatric patients compared to non-gated helical scans.
- The choice of scanning technique is crucial for minimizing radiation dose while maintaining diagnostic image quality.
- Adherence to the ALARA principle (As Low As Reasonably Achievable) is paramount in pediatric cardiac CTA.
Introduction:
Investigating the extent to which the exposure dose differs in cardiac computed tomography angiography (CTA) between electrocardiogram (ECG)-gated and non-gated helical scan is crucial. We aimed to investigate the difference in radiation dose with cardiac CTA between ECG-gated and non-gated helical scans in pediatric patients.
Methods:
Between January 2015 and November 2021, for the study cohort for the retrospective cardiac CTA study, we enrolled 63 pediatric patients aged 1-60 months and divided them into non-gated helical (33 patients) and retrospective ECG-gated (30 patients) scans with a 64-detector-row CT. We compared the CT dose-volume index (CTDIvol), dose-length product (DLP), effective dose (ED), CT values of the aorta and pulmonary artery, and image noise between the two groups. We calculated 95 % confidence intervals (CIs).
Results:
Compared with the non-gated helical and ECG gated scans, CTDIvol were 0.5 ± 0.1 and 5.7 ± 1.4 mGy (P < 0.01: 95 % CI, from -5.7 to -4.7), DLPs were 8.6 ± 2.8 and 72.6 ± 24.6 mGy・cm (P < 0.01: 95 % CI, from -72.6 to -55.3), and EDs were 0.2 ± 0.1 and 1.9 ± 0.6 mSv (P < 0.01: 95 % CI, from -1.9 to -1.4), respectively. Radiation dose for retrospective ECG-gated scans was approximately 10 times higher than that for non-gated helical scans. There were no significant differences in the CT values and image noise between the retrospective ECG-gated and non-gated helical scans (all p < 0.05).
Conclusion:
The radiation dose for retrospective ECG-gated scans with cardiac CTA was approximately 10 times higher than that for non-gated helical scans without any reduction in image noise or contrast enhancement.
Implications For Practice:
The radiation dose should be as low as reasonably achievable, and the selection of the scan method is pivotal in clinical practice.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System IV: CMRI