Related Experiment Video
Updated: Jun 14, 2025

Using Extraordinary Optical Transmission to Quantify Cardiac Biomarkers in Human Serum
Published on: December 13, 2017
Derivation and Validation of ESC-0/1-h Algorithm for High-Sensitivity Troponin T and I in Cancer Patients
Paolo Bima1, Pedro Lopez-Ayala2, Luca Koechlin3
1Cardiovascular Research Institute Basel (CRIB) and Department of Cardiology, University Hospital Basel, University of Basel, Basel, Switzerland; GREAT Network, Rome, Italy; Dipartimento di Scienze Mediche, Università degli Studi di Torino, Torino, Italy.
New European Society of Cardiology (ESC) 0/1-hour high-sensitivity cardiac troponin (hs-cTnT/I) algorithm cutoffs optimized for cancer patients improve non-ST-elevation myocardial infarction diagnosis efficacy and safety.
Area of Science:
- Cardiology
- Oncology
- Biomarker Diagnostics
Background:
- Diagnostic performance of high-sensitivity cardiac troponin T/I (hs-cTnT/I) and European Society of Cardiology (ESC) 0/1-h algorithms for non-ST-elevation myocardial infarction (NSTEMI) is reduced in cancer patients.
- Cancer patients present unique challenges for NSTEMI diagnosis due to potential biomarker alterations.
Purpose of the Study:
- To derive and validate novel hs-cTnT/I cutoffs within the ESC 0/1-h algorithm framework, specifically tailored for patients with active or past cancer.
- To enhance the diagnostic accuracy and clinical utility of hs-cTnT/I testing in the cancer population for NSTEMI detection.
Main Methods:
- International multicenter study analyzing patients with suspected NSTEMI.
- Central adjudication of final diagnoses by two independent cardiologists based on the fourth universal definition of myocardial infarction.
- External validation in two independent cohorts to confirm the generalizability of the derived cutoffs.
Main Results:
- Cancer-optimized ESC 0/1-h hs-cTnT cutoffs significantly increased diagnostic efficacy from 58.6% to 68.0% (P < 0.001), maintaining high sensitivity and specificity.
- Cancer-optimized ESC 0/1-h hs-cTnI-Architect cutoffs improved efficacy from 59.3% to 78.9% (P < 0.001), with preserved sensitivity and specificity.
- Findings were robustly validated in internal and external patient cohorts.
Conclusions:
- The developed cancer-optimized ESC 0/1-h hs-cTnT/I algorithm cutoffs enhance diagnostic efficacy for NSTEMI in cancer patients.
- These optimized cutoffs maintain high safety profiles, ensuring reliable rule-out and rule-in of myocardial infarction.