Related Experiment Video
Updated: Jun 12, 2025

16:40
A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
26.2K
Variability in implantable cardioverter-defibrillator battery longevity
James V Freeman1, Michael Torre2, Prashanthan Sanders3
1Department of Medicine, Yale University School of Medicine, New Haven, Connecticut.
Heart Rhythm
|May 24, 2025
Summary
Implantable cardioverter-defibrillator (ICD) battery longevity is influenced by device type and manufacturer. Newer ICDs show improved longevity, with manufacturer and programming settings being key factors.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Device Technology
Background:
- Implantable cardioverter-defibrillator (ICD) battery life directly affects generator replacement frequency, influencing patient risk and healthcare costs.
- Optimizing ICD battery longevity is crucial for reducing complications and managing economic burdens associated with device replacement.
Purpose of the Study:
- To identify key factors influencing the battery longevity of implantable cardioverter-defibrillators (ICDs).
- To compare the battery longevity performance across different ICD manufacturers.
Main Methods:
- Analysis of a nationwide, multicenter remote monitoring dataset (PaceMate) for ICDs implanted between 2003 and 2023.
- Comparison of Kaplan-Meier survival curves and device parameters across manufacturers, evaluating time to replacement interval (RI).
- Cox regression analysis to determine the impact of device type, manufacturer, and programmed parameters on battery longevity.
Main Results:
- Each additional lead decreased ICD battery longevity by approximately 2-3 years.
- Boston Scientific transvenous ICDs demonstrated the longest battery longevity across single-, dual-, and biventricular types.
- ICDs implanted in the last 2 years showed a 2-4 year improvement in estimated longevity compared to older devices.
Conclusions:
- ICD battery longevity significantly varies by device type and manufacturer, with newer devices offering substantially longer lifespans.
- Manufacturer, programmed pulse width, and programmed output are the primary determinants of ICD battery longevity.
- Understanding these factors can guide clinical decisions and improve patient outcomes by optimizing device selection and programming.
More Related Videos
Related Concept Videos
Cardiomyopathy V: Interprofessional Care
2
Managing cardiomyopathy involves addressing underlying or precipitating causes, treating heart failure with medications, and implementing dietary changes and a balanced exercise and rest regimen.Lifestyle ModificationsCardiomyopathy patients should adopt a low-sodium diet to reduce fluid retention and manage heart failure. A personalized exercise and rest plan helps maintain physical fitness without overstraining the heart. Avoiding alcohol and tobacco is essential to prevent further damage to...
2
Cardiomyopathy II: Dilated Cardiomyopathy
1
Dilated cardiomyopathy, or DCM, is a progressive myocardial disorder characterized by ventricular chamber dilation and contractile dysfunction.EtiologyVarious factors can cause DCM, including hypertension and heavy alcohol intake, which contribute to the weakening and enlargement of the heart muscle. Viral infections, such as Coxsackievirus B, adenoviruses, and influenza, can lead to DCM by causing inflammation and damage to heart tissue. Certain chemotherapeutic agents, including daunorubicin,...
1