A Retrospective Observational Study of Vibrating Mesh Nebulizers in the Pediatric Emergency Department
Caitlin E Crumm1,2, Robert M DiBlasi3,4,5, Dwight Barry3
1Division of Pediatric Emergency Medicine, Seattle Children's.
Objective:
To determine outcomes of bronchodilator delivery with a vibrating mesh nebulizer (VMN) in a pediatric emergency department (ED).
Methods:
We performed an in vitro analysis of the VMN and a retrospective observational study of patients receiving nebulized albuterol for a respiratory indication before (April 24, 2022-August 11, 2022) and after (April 24, 2023-July 29, 2023) introduction of the VMN in the ED.
Results:
Six hundred five children ages 1 to 17 years were included (302 preintervention, 303 postintervention). The adjusted total albuterol dose per patient was less in the postintervention group (-15.8 mg; 95% CI: -17.5 mg, -14.2 mg). The adjusted time to disposition was less post-intervention (-36 min; 95% CI: -51, -20). The adjusted odds of admission to the hospital or if admitted to the intensive care unit (ICU) were not statistically different [admission adjusted odds ratio (aOR): 1.1; 95% CI: 0.7, 1.7; ICU admission if admitted aOR: 1.7; 95% CI: 0.7, 4.3]. Adjusted mean change in respiratory score with first nebulizer was similar (-0.2; 95% CI: -0.6, 0.3). Within 72 hours of disposition, odds of unplanned transfer to the ICU after admission were similar (aOR: 0.4; 95% CI: 0.1, 1.6), as was unplanned ED return after discharge (aOR: 0.7; 95% CI: 0.3, 1.9).
Conclusions:
VMN use in the pediatric ED was associated with a lower total albuterol dose per patient as well as a shorter time to disposition. It did not change the odds of admission, admission to the ICU, unplanned transfers to the ICU after admission, or returns to the ER within 72 hours of discharge.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Suctioning the Nasopharyngeal Airway
Equipment Required
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Inhaled Medications
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...