Far-Red, High-Resolution, Reflection-Free Images of the Anterior Segment in Retro-Illumination
Anthony Ain1, Sylvain Poinard1, Thierry Lepine1,2,3
1Laboratory for Biology, Engineering and Imaging for Ophthalmology, BiiO, Faculty of Medicine, Health & Innovation Campus, Jean Monnet University, Saint-Etienne, France.
Purpose:
Examination of the anterior segment using retro-illumination is part of the routine clinical examination, but capturing images that can be used by the clinician remains a challenge because reflections are always present, resolution is insufficient, and the patient's glare often causes motion blur. Our aim was therefore to overcome these limitations.
Methods:
We modified a slit lamp by using a far-red LED, a high-resolution camera, and by modifying the light path using polarizers and an obturator to suppress reflections while optimizing the light flux. We used this prototype to image various ocular conditions to illustrate its potential.
Results:
With the ×40 magnification of the slit lamp, the prototype provided images with a resolution of 6.2 µm, and a field of view of 9.7 mm. We obtained images without motion blur in all patients. The vast majority of images were reflection-free. Only the images of a few patients with pseudophakia showed persistent small reflection. The resolution was sufficient to highlight, for example, all the Descemetic excrescences (Guttae) constitutive of a Fuchs endothelial corneal dystrophy (FECD).
Conclusions:
Our retro-illumination device prototype provides high-resolution images of the different structures of the anterior segment of the eye, easy to acquire mainly because they do not dazzle patients. We believe that these images have the potential to facilitate the diagnosis and monitoring of many different diseases both in routine use and in clinical trials where they could be used as objective endpoints.
Translational Relevance:
Our prototype revisits retro-illumination with a potential to become a new imaging device.
More Related Videos
10:14Author Spotlight: Ex Vivo OCT-Based Multimodal Imaging of Human Donor Eyes for Research into Age-Related Macular Degeneration
Published on: May 26, 2023
06:19In Vivo Imaging of Cx3cr1gfp/gfp Reporter Mice with Spectral-domain Optical Coherence Tomography and Scanning Laser Ophthalmoscopy
Published on: November 11, 2017
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Total Internal Reflection Fluorescence Microscopy
Super-resolution Fluorescence Microscopy
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...