Diffusion MRI in the cortex of the brain: Reducing partial volume effects from CSF and white matter in the mean diffusivity using high b-values and spherical b-tensor encoding
Related Experiment Video
Updated: Sep 19, 2025

Diffusion Imaging in the Rat Cervical Spinal Cord
Published on: April 7, 2015
Cornelia Säll1, Nicola Spotorno2, Pia C Sundgren3,4,5
1Department of Medical Radiation Physics, Lund University, Lund, Sweden.
Purpose:
The mean diffusivity (MD) is sensitive to the microstructure of the cortex. However, partial volume effects with CSF and white matter (WM) may obscure pathology-related alterations. This work investigates both existing approaches and a novel approach for reducing partial volume effects.
Theory And Methods:
A bias in MD arises due to partial volume effects, higher-order terms, and the noise floor in magnitude data. We propose to reduce this bias by using high b-value encoding to limit partial volume effects with CSF, spherical b-tensor encoding to reduce the influence of higher-order terms, and super-resolution acquisition and reconstruction to suppress the noise floor. This approach was investigated, along with established approaches (inversion recovery and free water elimination) for reducing partial volume effects, using simulations and in vivo data.
Results:
High b-value diffusion MRI with spherical b-tensor encoding reduced partial volume effects with CSF relative to conventional diffusion MRI. Maximum errors decreased from 0.51 to 0.01 μm2/ms in simulations. In vivo, the median absolute deviation of cortical MD decreased from 0.17 to 0.06 μm2/ms, whereas the median decreased slightly from 0.77 to 0.73 μm2/ms. The other methods yielded bias from either CSF, WM, or model assumptions.
Conclusion:
The mean diffusivity of the cortex can be mapped in high precision with reduced influence of partial volume effects with CSF and WM matter using high b-values and spherical b-tensor encoding and super-resolution reconstruction.
Insights
This study introduces a novel diffusion MRI method to accurately measure cortical mean diffusivity by minimizing partial volume effects. The new technique significantly improves precision, offering clearer insights into brain microstructure.
Area of Science:
- Neuroimaging
- Diffusion MRI
Background:
- Mean diffusivity (MD) reflects cortical microstructure but is affected by partial volume effects with cerebrospinal fluid (CSF) and white matter (WM).
- Existing methods for reducing partial volume effects have limitations and can introduce bias.
Purpose of the Study:
- To investigate existing and novel approaches for reducing partial volume effects in cortical MD measurements.
- To develop a more precise method for mapping cortical MD, reducing bias from CSF and WM.
Main Methods:
- Proposed a novel approach using high b-value diffusion MRI with spherical b-tensor encoding and super-resolution reconstruction.
- Investigated the novel approach alongside established methods like inversion recovery and free water elimination.
- Validated the methods using both simulations and in vivo data.
Main Results:
- The novel approach significantly reduced partial volume effects with CSF compared to conventional MRI, decreasing maximum errors in simulations from 0.51 to 0.01 μm²/ms.
- In vivo, the median absolute deviation of cortical MD decreased from 0.17 to 0.06 μm²/ms.
- Established methods introduced bias from CSF, WM, or model assumptions, while the novel method showed superior performance.
Conclusions:
- High b-value diffusion MRI with spherical b-tensor encoding and super-resolution reconstruction enables high-precision mapping of cortical MD.
- This advanced technique effectively reduces the influence of partial volume effects with CSF and WM.
- The findings offer a more accurate way to assess cortical microstructure and potential pathology.