Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
Novel Fluoroscopic Evaluation of Incorrect Impella Placement.
Andrew P Hill1, Michael Slack2, Ron Waksman3
1Department of Cardiology, MedStar Washington Hospital Center, Washington, DC, USA.
A novel fluoroscopic technique identified incorrect Impella placement, preventing complications during percutaneous coronary intervention for cardiomyopathy and coronary artery disease.
Area of Science:
- Cardiology
- Interventional Cardiology
- Medical Imaging
Background:
- Percutaneous coronary intervention (PCI) is a key treatment for coronary artery disease.
- The Impella device is used to support patients during high-risk PCI, especially those with cardiomyopathy.
- Ensuring correct Impella placement is crucial for device efficacy and patient safety.
Purpose of the Study:
- To describe a novel fluoroscopic technique for assessing Impella catheter placement.
- To highlight potential complications arising from malrotation of the Impella device.
Main Methods:
- A case study of a 65-year-old man with cardiomyopathy and multivessel coronary artery disease undergoing Impella-supported PCI.
- Utilized standard single-plane fluoroscopy and additional non-standard fluoroscopic views to evaluate Impella catheter positioning.
- Identified malrotation of the Impella catheter toward the mitral apparatus.
Main Results:
- Initial assessment with standard fluoroscopy suggested correct Impella placement.
- Additional fluoroscopic views revealed malrotation of the Impella catheter.
- This malrotation was oriented toward the mitral apparatus, posing a risk for complications.
Conclusions:
- Malrotation of the Impella microaxial flow pump catheter can lead to serious complications and suboptimal hemodynamic support.
- This case demonstrates the utility of a novel fluoroscopic technique in accurately evaluating Impella placement.
- This technique can aid in preventing complications associated with incorrect Impella positioning during PCI.