Imageless vs Image-Based Computer Navigation in Total Hip Arthroplasty: An Intracase Comparison Shows Variability
Joseph T Gibian1, Ryan J Cone1,2, Matthew Booth1
1Department of Orthopaedics, Washington University School of Medicine, St. Louis, MO, USA.
Computer navigation in total hip arthroplasty shows significant measurement discordance between imageless (ILCN) and image-based (IBCN) systems. Factors like BMI and prior spine fusion increase anteversion measurement discrepancies.
Area of Science:
- Orthopedic Surgery
- Medical Technology
- Surgical Navigation
Background:
- Computer navigation is increasingly used in total hip arthroplasty (THA).
- Two main modalities exist: imageless (ILCN) and image-based (IBCN).
- Comparing these navigation systems is crucial for accurate component placement.
Purpose of the Study:
- To compare inclination and anteversion measurements between ILCN and IBCN within the same THA case.
- To identify the frequency of discordant measurements between the two navigation modalities.
- To investigate factors associated with measurement discordance.
Main Methods:
- One hundred consecutive THAs utilized both ILCN and IBCN.
- Intraoperative acetabular component inclination and anteversion were recorded for both systems.
- Intracase measurements were compared, and discordant cases (>5° difference) were identified.
- Linear regression analyzed confounding factors for discordance.
Main Results:
- No significant difference in mean inclination or anteversion between ILCN and IBCN.
- Significant discordance observed: 21% for inclination and 32% for anteversion.
- Increased body mass index and prior lumbar spine fusion correlated with anteversion discordance.
- No association found between discordance and dislocation rates.
Conclusions:
- Over 20% of THA cases exhibit discordant measurements between ILCN and IBCN.
- Lumbar spine fusion and elevated body mass index are risk factors for anteversion measurement discordance.
- These findings highlight potential variability in computer-assisted THA outcomes.