Background:
Paediatric patients with congenital heart disease benefit greatly from X-ray diagnostic and interventional procedures. However, multiple procedures lead to prolonged exposure to ionizing radiation (IR), raising concerns for secondary long-term health issues.
Purpose:
This study aimed to establish local dose reference levels (LDRLs) for a National Reference Center for Congenital Cardiopathies (NRCCC) in Portugal. Additionally, it estimated effective and organ doses in paediatric patients undergoing coronary angiographies (CAs) using measurements and Monte Carlo simulations.
Methods:
LDRLs were determined through statistical analysis of patient registry data from 120 patients recorded between 2022 and 2023. For effective dose estimation, personal dose equivalent values Hp(10) were measured with an anthropomorphic phantom (CIRS ATOM 705) and Raysafe dosimeters, used to validate Monte Carlo (MC) simulations, which were then employed with rescaled paediatric voxel phantoms (GsF Baby and Child), to estimate effective doses and dose conversion coefficients (DCCs) for representative patient anatomies across different age and weight groups.
Results:
LDRLs for coronary angiographies at the NRCCC showed significant dose variability, indicating dose optimization opportunities. Monte Carlo simulations allowed for organ dose and effective dose calculations. Dose Conversion Coefficients were 1.107, 0.882, 0.719, and 0.524 mSv/Gy·cm2 for newborn, 1 year-old, 5 year-old and 10 year-old groups respectively. A strong linear correlation between effective dose and air-kerma area product (KAP) was observed.
Conclusions:
The study highlights the strong correlation between effective dose and KAP, offering a practical framework for estimating patient doses and enhancing radiation safety protocols in clinical practice.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Dosage Regimens: Designs and Approaches
Therapeutic Drug Monitoring: Drug Analysis Methods
Drug Dosing: Infants and Children
Drug Dosing in Renal Diseases: Measurement of Glomerular Filtration Rate
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration