Ventricular arrhythmia detection with a wearable ring-type photoplethysmography sensor: A feasibility study
Nicole J van Steijn1, Shari Pepplinkhuizen1, Pieter G Postema1
1Department of Cardiology, Heart Centre, Amsterdam University Medical Centre, Location Academic Medical Centre, Amsterdam, The Netherlands.
Heart Rhythm
|June 15, 2025
Summary
A novel ring-type wearable photoplethysmography (PPG) sensor shows high sensitivity in detecting ventricular arrhythmias (VAs), including ventricular fibrillation and tachycardia. This advancement offers potential for early detection of sudden cardiac death triggers.
Area of Science:
- Cardiology
- Biomedical Engineering
- Wearable Technology
Background:
- Ventricular arrhythmias (VAs) are a primary cause of sudden cardiac death.
- Early VA detection is critical for patient survival.
- Wrist-worn photoplethysmography (PPG) sensors show limited success in VA detection; ring-type sensors remain uninvestigated.
Purpose of the Study:
- To evaluate the sensitivity of a ring-type wearable PPG sensor in detecting induced VAs.
- To compare PPG sensor data with electrocardiogram (ECG) for VA detection accuracy.
Main Methods:
- Patients undergoing VA induction procedures were enrolled.
- A ring-type PPG sensor (CART-I) and ECG were used for data collection.
- Sensitivity was calculated, and Bland-Altman analysis assessed agreement in arrhythmia duration.
Main Results:
- The study included 25 patients with 67 VA episodes (27 ventricular fibrillation, 40 ventricular tachycardia).
- The ring-type PPG sensor achieved 94% sensitivity in detecting VAs (100% for VF, 90% for VT).
- Excellent agreement (ICC 0.998) was found between PPG and ECG for arrhythmia duration.
Conclusions:
- A ring-type wearable PPG device demonstrates high sensitivity for detecting VAs.
- This technology shows promise for non-invasive, early detection of life-threatening arrhythmias.
Related Concept Videos
Pulse rhythm
940
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
940
Dysrhythmias V: Evaluating Dysrhythmias
123
Dysrhythmias, also known as arrhythmias, are disturbances in the heart's rhythm that range from benign to life-threatening. A thorough evaluation is crucial for appropriate management and involves a comprehensive medical history, physical examination, and various diagnostic tests.Medical HistorySymptoms: Collect detailed information on palpitations, dizziness, syncope, chest pain, and fatigue. Note their onset, frequency, and triggers.Previous Cardiac Issues: Document any history of heart...
123