Systematic Quality Assurance/Quality Control Framework for a Volatile Organic Compound Real-Time Monitoring
Hong Cheng Tay1,2, Anthony Miller3, Hunter N B Moseley2,4
1Department of Civil Engineering, University of Kentucky, Lexington, Kentucky 40506, United States.
None:
This study presents a systematic quality assurance/quality control (QA/QC) framework for new technology adopters to validate environmental monitoring instruments when consistent, standardized validation and calibration procedures do not readily exist. This QA/QC framework is especially relevant for newly developed environmental monitoring instruments prior to deployment in field settings as part of environmental site investigations. Key QA/QC parameters used in the framework include repeatability, intermediate precision, reproducibility, linearity, limits of detection (LOD), limit of quantification (LOQ), trueness, and recovery. The QA/QC framework is demonstrated on the autonomous rugged optical multigas analyzer for VOCs (AROMA-VOC), a technology that employs cavity ring-down spectroscopy (CRDS) with preconcentration and chromatographic separation prior to CRDS detection. QA/QC protocols for nine VOCs, and results for AROMA-VOC were compared with laboratory-based gas chromatography/mass spectrometer (GC-MS). Results demonstrated high linearity (R 2 > 0.95) and low LOQ values (0.0004 ± 0.0002-0.0455 ± 0.0192 μg/L). The instrument also exhibited ranges of LOD at least an order of magnitude lower than that of the referenced GC-MS. AROMA-VOC demonstrated acceptable precision [repeatability, intraday relative standard deviation (RSD %): 1.72-12.1%; intermediate, interday RSD %: 2.01-10.93%] and substantial recovery (104.8% to 212.4%). Interlaboratory testing also concluded that AROMA-VOC can produce consistent detections (R 2 > 0.90). The z-score findings suggested that the performance of the instrument was satisfactory, with the exceptions of xylenes and 1,3-butadiene. Overall, using the systematic QA/QC framework presented in this study, AROMA-VOC is shown to be a robust method for volatile organic compound analysis.
More Related Videos
07:57Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
10:22Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements
Published on: September 7, 2019
Related Concept Videos
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...