Consensus recommendations for hyperpolarized [1-13C]pyruvate MRI multi-center human studies.
Shonit Punwani1,2, Peder E Z Larson3, Christoffer Laustsen4
1Department of Radiology, University College London Hospital NHS Foundation Trust, London, UK.
Magnetic Resonance in Medicine
|June 16, 2025
Summary
Consensus was reached on best practices for multi-center hyperpolarized (HP) [1-13C]pyruvate MRI studies. This advances the clinical application of this in vivo metabolism assessment technology.
Area of Science:
- Medical Imaging
- Biophysics
- Metabolic Research
Background:
- Hyperpolarized (HP) [1-13C]pyruvate MRI enables in vivo assessment of cellular metabolism.
- This technique has been translated to human studies across various diseases at multiple international centers.
Purpose of the Study:
- To establish consensus on best practices for conducting multi-center human studies using HP [1-13C]pyruvate MRI.
- To guide the development and standardization of clinical applications for this imaging modality.
Main Methods:
- A two-round formal consensus-building exercise involving 29 experts from 13 sites.
- Experts represented pharmacy methods, MR physics, translational imaging, and data analysis.
- 246 statements covering preparation, system setup, acquisition, reconstruction, and data analysis were evaluated.
Main Results:
- Consensus was achieved on approximately two-thirds of the statements across all evaluated categories.
- Key agreements include standardized release criteria for HP pyruvate preparation and phantom-based quality assurance for site qualification.
- Preferences were noted for specific pulse sequences and reconstruction methods based on anatomical targets, and for normalized metabolite area-under-curve values as preferred metrics.
Conclusions:
- HP [1-13C]pyruvate MRI technology is sufficiently mature for planning multi-center studies.
- The consensus process identified areas of agreement for multi-center study conduct and highlighted areas needing further research to ascertain optimal best practices.