B1-corrected breast T1 mapping at ultralow field.
Sheng Shen1,2, Neha Koonjoo1,2, Stephen E Ogier3,4
1A.A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Boston, Massachusetts, USA.
Magnetic Resonance in Medicine
|June 16, 2025
Summary
This study introduces an efficient ultralow-field (ULF) MRI technique for accurate T1 mapping of breast tissue. The method corrects for B1 inhomogeneity, providing reliable T1 values without fat suppression.
Area of Science:
- Medical Imaging
- Magnetic Resonance Imaging
- Biophysics
Background:
- Accurate T1 mapping is crucial for quantitative MRI in various tissues.
- B1 field inhomogeneity poses a significant challenge for T1 mapping accuracy, especially in ultralow-field (ULF) MRI.
- Developing bias-free T1 mapping methods is essential for reliable diagnostic applications.
Purpose of the Study:
- To develop an efficient T1 mapping approach for ultralow-field (ULF) MRI that corrects for B1 inhomogeneity.
- To determine the T1 values of breast tissues at 6.5 mT without fat suppression.
- To validate the developed T1 mapping method in vitro and in vivo.
Main Methods:
- A variable flip angle approach was employed for simultaneous T1 and B1 mapping.
- B1 maps were used to correct T1 maps, improving accuracy.
- The method was validated using copper sulphate phantoms and a breast phantom, followed by in vivo scans of healthy volunteers.
Main Results:
- T1 map measurements in phantoms showed less than 13% deviation from reference T1 values.
- For reference T1 values greater than 50 ms, the deviation was less than 6%.
- In vivo T1 maps of breast tissue in healthy volunteers at 6.5 mT showed 25th to 75th percentiles ranging from 79 to 243 ms.
Conclusions:
- The variable flip angle method is effective for simultaneous B1 and T1 mapping at ULF.
- This approach enables efficient and accurate in vivo T1 mapping in ULF MRI.
- The study provides reference T1 values for breast tissue at 6.5 mT.