Using Durometers to Quantify Stiffness as an Outcome Measure for Cutaneous Neurofibroma in Neurofibromatosis Type 1
Bahir Chamseddin1, Renee M McKay2, Ruyun Jin3
1Department of Dermatology, University of Texas Southwestern Medical Center, Dallas, Texas, USA.
Summary
This study found that both the Rex Gauge durometer (REX) and Delfin SkinFibroMeter (DELFIN) reliably measure cutaneous neurofibroma (cNFs) stiffness. These instruments offer consistent quantification for clinical trials evaluating new neurofibromatosis type 1 treatments.
Area of Science:
- Biomedical Engineering
- Dermatology
- Oncology
Background:
- Neurofibromatosis type 1 (NF1) patients develop numerous cutaneous neurofibromas (cNFs).
- Currently, no effective drug therapies exist for cNFs; physical removal is the primary treatment.
- Quantitative methods are needed to assess treatment efficacy in clinical trials, particularly changes in tumor stiffness.
Purpose of the Study:
- To evaluate the reproducibility of two instruments, the Rex Gauge durometer (REX) and Delfin SkinFibroMeter (DELFIN), in measuring cNF stiffness.
- To establish reliable quantitative techniques for assessing cNF changes in response to potential therapies.
Main Methods:
- Ninety-seven cNFs from NF1 patients were measured using both REX and DELFIN devices.
- Measurements were taken at two separate visits, approximately two weeks apart, to assess within-tumor and within-visit agreement.
- Intraclass correlation coefficients (ICCs) were calculated to determine measurement reliability, accounting for repeated measures and patient clustering.
Main Results:
- The REX demonstrated excellent within-visit agreement (ICC = 0.937) and moderate within-tumor agreement (ICC = 0.740).
- The DELFIN showed moderate within-visit agreement (ICC = 0.732) and moderate within-tumor agreement (ICC = 0.607).
- Both instruments were found to be user-friendly and provided consistent quantification of cNF stiffness.
Conclusions:
- Both the REX Gauge and Delfin SkinFibroMeter are reliable tools for quantifying cutaneous neurofibroma stiffness.
- These instruments can support clinical trials by providing objective measures of treatment response in neurofibromatosis type 1.
- Further research can utilize these validated tools to assess the efficacy of emerging therapies for cNFs.