In vivo Testing of the Pressio Intracranial Pressure Monitor
Laurent Gergelé1, Jérémy Mallard2, Clément Magand2
1Pôle d'anesthésie Réanimatuon, Service de Réanimation Neurochirurgicale, CHU de Clermont-Ferrand, Université de Clermont-Ferrand, Clermont-Ferrand, France. laurentgergele@yahoo.fr.
Neurocritical Care
|June 20, 2025
Summary
The Pressio intraparenchymal strain gauge transducer shows minimal zero drift in clinical settings, making it reliable for monitoring intracranial pressure in acute brain injury patients. Its precision is comparable to other strain gauge devices.
Area of Science:
- Neurosurgery
- Critical Care Medicine
- Biomedical Engineering
Background:
- Accurate intracranial pressure (ICP) monitoring is vital for managing acute brain injuries.
- The Pressio intraparenchymal strain gauge transducer is a device used for ICP measurement.
Purpose of the Study:
- To evaluate the zero drift of the Pressio intraparenchymal strain gauge transducer under clinical conditions.
- To assess the reliability and precision of the Pressio transducer for long-term ICP monitoring.
Main Methods:
- A prospective, observational, multicenter study was conducted in four French intensive care units.
- 235 patients with acute brain injuries requiring ICP measurement with the Pressio transducer were included.
- Zero drift was measured at sensor explantation, and ICP-related adverse events were collected.
Main Results:
- The median zero drift was 1 mm Hg (IQR 1-3 mm Hg) over a median monitoring period of 8 days (2,180 total monitoring days).
- A weak, clinically insignificant correlation was found between monitoring duration and zero drift (ρ=0.141, P=0.0357).
- 10% of transducers showed zero drift >5 mm Hg; 1.8% had ICP-related hematomas with no clinical impact; 2.6% experienced monitoring failures.
Conclusions:
- The Pressio transducer demonstrates minor zero drift, comparable to other strain gauge ICP devices.
- The device is precise and reliable for clinical intracranial pressure monitoring in patients with acute brain injuries.