An effective time reduction strategy in whole-body SPECT/CT studies using novel acquire during step mode without
Timothy Melhuish1, Amit Nautiyal2, Sofia Michopoulou2
1Department of Medical Physics, University Hospital Southampton, Southampton SO16 6YD, UK.
Acquire During Step (ADS) technology reduces whole-body SPECT/CT scan times by 15% without impacting diagnostic accuracy. This advancement in SPECT/CT imaging speeds up scans, potentially reducing motion artifacts and improving disease diagnosis.
Area of Science:
- Nuclear Medicine
- Medical Imaging
- Radiochemistry
Background:
- Whole-body SPECT/CT imaging is crucial for diagnosing metastatic disease and enabling dosimetry.
- Impractically long acquisition times limit the routine clinical use of whole-body SPECT/CT.
- Acquire During Step (ADS) technology offers a potential solution to reduce scan durations.
Purpose of the Study:
- To evaluate the efficacy of ADS technology in reducing whole-body SPECT/CT acquisition times.
- To determine if reduced acquisition times compromise the diagnostic quality and quantitative accuracy of SPECT/CT scans.
- To assess the impact of ADS on quantitative accuracy using phantom studies with 99mTc and 177Lu.
Main Methods:
- Phantom studies were conducted using 99mTc and 177Lu isotopes.
- SPECT acquisition modes, including step-and-shoot (SSM) and ADS, were compared at full, half, and quarter times.
- Quantitative accuracy was assessed by calculating recovery coefficients and SUVs using the xSPECT quantification tool.
Main Results:
- ADS mode reduced SPECT scan time by 15% compared to SSM, with faster acquisition per bed.
- No statistically significant differences in recovery coefficients were observed between SSM and ADS modes across different acquisition times.
- Background SUV measurements showed no significant changes between SSM and ADS modes for both isotopes.
Conclusions:
- ADS technology can decrease SPECT/CT scan time by 15% without compromising quantitative accuracy.
- Further time reductions are achievable with half-time or quarter-time ADS, enhancing routine practice efficiency.
- Faster WB SPECT acquisition improves patient comfort by reducing motion artifacts and enhances diagnostic accuracy and personalized dosimetry.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT