Related Experiment Videos
Evaluation of the BIOGRAM antimicrobial susceptibility test system
Journal of Clinical Microbiology
|November 1, 1985
Summary
The BIOGRAM system accurately determines antimicrobial susceptibility testing (AST) results. It showed 95.9% correlation with the reference broth microdilution method for common clinical bacterial isolates.
Area of Science:
- Clinical Microbiology
- Infectious Diseases
- Diagnostic Technology
Background:
- Accurate antimicrobial susceptibility testing (AST) is crucial for guiding effective antibiotic treatment.
- Disk diffusion methods offer a standardized approach to AST, but interpretation can be complex.
- The BIOGRAM system aims to automate MIC interpretation from disk diffusion zone diameters.
Purpose of the Study:
- To evaluate the performance of the BIOGRAM system for determining Minimum Inhibitory Concentrations (MICs).
- To compare BIOGRAM results against the established broth microdilution reference method.
- To assess BIOGRAM's accuracy with a diverse range of clinically relevant bacterial pathogens.
Main Methods:
- The BIOGRAM system was challenged with 511 clinical isolates.
- Isolates included Enterobacteriaceae, nonfermentative gram-negative bacteria, staphylococci, and enterococci.
- Results were compared to broth microdilution, with 10,085 organism-drug combinations evaluated.
Main Results:
- BIOGRAM demonstrated high overall agreement with the broth microdilution method.
- An overall correlation of 95.9% was achieved.
- Appropriate control organisms were used throughout the testing series.
Conclusions:
- The BIOGRAM system provides accurate MIC determination from disk diffusion zone diameters.
- BIOGRAM is a reliable tool for antimicrobial susceptibility testing in clinical microbiology.
- The system shows excellent correlation with the reference broth microdilution method, supporting its clinical utility.