An FSM-Assisted High-Accuracy Autonomous Magnetic Compensation Optimization Method for Dual-Channel SERF Magnetometers Used in Weak Biomagnetic Signal Measurement
Related Experiment Video
Updated: Sep 18, 2025

Optimized Setup and Protocol for Magnetic Domain Imaging with In Situ Hysteresis Measurement
Published on: November 7, 2017
Xinran Tian1, Bo Bao1, Ridong Wang1
1State Key Laboratory of Precision Measurement Technology and Instruments, Tianjin University, Tianjin 300072, China.
Abstract:
Atomic magnetometers based on the spin-exchange relaxation-free (SERF) regime have broad applications in bio-magnetic measurement due to their high sensitivity and miniaturized size. In this paper, we propose a SERF-based magnetometer using 1 × 2 polarization-maintaining fiber (PMF) with single-beam parameter optimization. The impacts of temperature, pumping laser power, and modulation amplitude on the magnetometer's response signal at the SERF regime are examined. Moreover, through the simulation of zero-field resonance, the compensation accuracy is optimized. To further improve the compensation stability and accuracy, a novel finite state machine (FSM)-assisted iterative optimization magnetic field compensation algorithm is proposed. A pT-level compensation resolution with an error below 1.6% is achieved, which lays the foundation for the subsequent application of biomagnetic measurement arrays.
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
Magnetostatic Boundary Conditions
Magnetic Susceptibility and Permeability
When diamagnetic materials are placed under an external magnetic field, the moments opposite to the field are induced. Hence, the susceptibility for diamagnets has a minimal negative value of 10-5–10-6. Since...