Related Experiment Video
Updated: Sep 18, 2025

Direct Detection of the Acetate-forming Activity of the Enzyme Acetate Kinase
Published on: December 19, 2011
[Determination of Acetate Content in Hemodialysis Solutions and Dialysis Concentrates by HPLC]
Huadong Wang1, Yue Wang2, Dong Liu3
1Henan Center for Drug Evaluation and Inspection (Henan Center for Vaccine Inspection), Zhengzhou, 450000.
Abstract:
This study establishes a high-performance liquid chromatography (HPLC) method for the determination of acetate content in hemodialysis solutions and dialysis concentrates. In this study, Synergi Polar-RP column is utilized. Phosphate buffered saline (50 mmol/L, pH=2.5) is used as a mobile phase. The flow rate is 1.0 mL/min. The wavelength of detection is 212 nm. Results show that the linear relationship of acetate is good in the range of 0.1~20 mmol/L, r =0.999 9 and the spike recoveries are from 98.9%~99.5%, RSD<0.5% ( n=3). This method can easily and accurately determine the acetate content in hemodialysis solutions and dialysis concentrates, and can be applied to quality control in the production and use of such products.