Hydrostatic and central venous pressure measurement by the IVAC 560 infusion pump
Abstract:
The IVAC Variable Pressure Volumetric Pump, Model 560, was evaluated for hydrostatic and central venous pressure (CVP) measurement. In bench tests, 4 per cent of IVAC systems failed a +/- 2 mm Hg accuracy test. When measurements were inaccurate, errors were small (SD = 0.49 mm Hg). In simultaneous clinical measurements with a standard electronic transducer system, the correlation coefficient between the IVAC and transducer measurements was 0.95. A potential drawback of the IVAC 560 for CVP measurement is the lack of waveform display. The IVAC system is simple to operate and less expensive per use than the standard electronic system.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Measurement of Blood Pressure
Venous Return
What is Venous Return?
Venous return refers to the rate at which blood flows back to the heart from the body's peripheral veins. It's an integral part of the circulatory system as it...
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...