Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Machine-learning-based integration of temporal and spectral prompt gamma-ray information for proton range
Aaron Kieslich1,2, Sonja M Schellhammer1,2,3, Alex Zwanenburg1,4,5
1OncoRay - National Center for Radiation Research in Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, TUD Dresden University of Technology, Helmholtz-Zentrum Dresden-Rossendorf, Dresden, Germany.
Background And Purpose:
Prompt gamma-ray timing (PGT) and prompt gamma-ray spectroscopy (PGS) are non-invasive techniques for dose delivery monitoring in proton radiotherapy. Integrating PGT and PGS into a unified data analysis framework may improve proton range verification by incorporating both temporal and spectral information from prompt gamma-ray events. This study evaluates the effectiveness of this integration for enhancing the accuracy of proton range verification using machine-learning.
Material And Methods:
A homogeneous phantom was irradiated with 162 and 225 MeV static and scanned proton beams. Air cavities of 5, 10 and 20 mm were introduced to simulate anatomical variations. The energy and time of arrival of prompt gamma rays were measured using a PGT detector. 2-dimensional time-energy spectra were extracted for 1,440 proton spots. Different feature sets (energy-only, time-only, energy-restricted time, image) were computed. These feature sets were used by four different machine-learning models to predict range shifts. Model performance was assessed using the root mean square error (RMSE).
Results:
Time-only and combined time-energy feature sets exhibited good performance with RMSE values of 3 to 4 mm, consistent with previously developed models. Energy-only and image features led to poorer performance with RMSE values exceeding 5 mm. The integration of energy-only features did not improve prediction accuracy compared to exclusively using time-only features.
Conclusion:
While spectral information did not contribute additional value for determining proton beam range shifts in the investigated setup, the findings show that temporal information alone is sufficient to perform accurate proton range verification.
More Related Videos
10:03Proton Transfer and Protein Conformation Dynamics in Photosensitive Proteins by Time-resolved Step-scan Fourier-transform Infrared Spectroscopy
Published on: June 27, 2014
07:24Paramagnetic Relaxation Enhancement for Detecting and Characterizing Self-Associations of Intrinsically Disordered Proteins
Published on: September 23, 2021
Related Concept Videos
¹H NMR of Labile Protons: Temporal Resolution
The –OH proton in alcohols typically appears in the range of δ 2 to 5 ppm but can vary depending on the specific...
Atomic Emission Spectroscopy: Instrumentation
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...