Related Experiment Video
Updated: Oct 13, 2026

Stereotactic Radiosurgery for Gynecologic Cancer
Published on: April 17, 2012
Target-volume-based prescription selection for specified adverse-effect levels in automated stereotactic radiosurgery
Joel A Pogue1, Rex Cardan1, Farnaz Rahim Li1
1Department of Radiation Oncology, The University of Alabama at Birmingham, 619 19th Street South, Birmingham, AL 35249, USA.
Abstract:
Automated stereotactic radiosurgery (SRS) planning produced consistent dose falloff, enabling adverse-effect-relevant isodose volumes ( ) to be estimated before plan generation. We derived a direct isoeffect prescription dose ( ) model from previously validated target-volume ( ) power laws for single-isocenter automated SRS. coefficients and exponents were modelled as continuous functions of relative isodose level ( ), allowing to be solved from , desired , and HyTEC brain adverse-effect dose thresholds. Across held-out levels, target-level residuals had a median of 0.2%, an interquartile range of [-5.2%, +7.0%], and a 95% interval of [-12.8%, +27.0%].