Related Experiment Video
Updated: Aug 17, 2026

Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Impact of patient size on image quality of OSEM3D and BSREM reconstructions in [68Ga]Ga-DOTA-TATE PET/MR
Christina P W Cox1, Tessa Brabander2, Frederik A Verburg2
1Department of Radiology & Nuclear Medicine, Erasmus Medical Centre, Postbus 2040, 3000 CA, Rotterdam, The Netherlands. c.cox@erasmusmc.nl.
Background:
Previous [68Ga]Ga-DOTA-TATE PET/CT studies using ordered subset expectation maximization (OSEM3D) based reconstruction algorithms, demonstrated non-linear relations between body weight and image quality. Block Sequential Regularized Expectation Maximization (BSREM) algorithm reduces noise amplification during reconstruction. The impact of the reconstruction algorithm on the relation between image quality and patient size in [68Ga]Ga-DOTA-TATE PET/MR may differ from PET/CT and OSEM3D. Therefore, the aim of this study is to investigate the relation between patient size and image quality in OSEM3D and BSREM [68Ga]Ga-DOTA-TATE PET/MR reconstructions.
Methods:
[68Ga]Ga-DOTA-TATE PET/MR images of 55 patients were included. Images were reconstructed using OSEM3D (VUE Point FX SharpIR, 4 iterations, 28 subsets and 7 mm Gaussian filter) and BSREM (Q.Clear, β = 300). Liver signal-to-noise ratio (SNRliver) normalized for injected activity and acquisition time (SNRliver,norm) was measured to perform curve fitting with patient-dependent parameters using fixed, linear and non-linear fit models, followed by Akaike's corrected information criterion (AICc) model selection.
Results:
BSREM mean SNRliver was significantly (p < 0.001) higher than OSEM3D mean SNRliver. Body mass, the best patient-dependent parameter for both algorithms, clarified 40% (linear model) and 53% (non-linear model) of the variability in SNRliver,norm for OSEM3D and 20% (linear model) and 21% (non-linear model) for BSREM. AICc preferred a non-linear model for OSEM3D and a linear model for BSREM.
Conclusion:
The image quality predictor body weight is a weaker predictor for BSREM than for OSEM3D image quality in [68Ga]Ga-DOTA-TATE PET/MR. Therefore, a linear dosage regimen based on body weight is preferable for BSREM, whereas a quadratic dosage regimen based on body weight is optimal for OSEM3D.
More Related Videos
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024