Photon-Counting Detector CT of the Brain Reduces Variability of Hounsfield Units and Has a Mean Offset Compared with
Thomas Stein1, Friederike Lang1, Stephan Rau1
1From the Department of Diagnostic and Interventional Radiology (T.S., F.L., S.R., M.R., M.F.R., T.S., A.F., J.W., F.B.), Medical Center-University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany.
Background And Purpose:
Distinguishing GM from WM is essential for CT of the brain. The recently established photon-counting detector (PCD)-CT technology uses a novel detection technique that might allow more precise measurement of tissue attenuation for an improved delineation of attenuation values (Hounsfield units [HU]) and improved image quality in comparison with energy-integrating detector (EID)-CT. To investigate this, we compared HU, GM versus WM contrast, and image noise by using automated deep learning-based brain segmentations.
Materials And Methods:
We retrospectively included patients who received either PCD-CT or EID-CT and did not display a cerebral pathology. A deep learning-based segmentation of the GM and WM was used to extract HU. From this, the gray-white matter ratio (GWR) and contrast-to-noise ratio (CNR) were calculated.
Results:
We included 329 patients with EID-CT (mean age 59.8 ± 20.2 years) and 180 with PCD-CT (mean age 64.7 ± 16.5 years). GM and WM showed significantly lower HU in PCD-CT (GM: 40.4 ± 2.2 HU; WM: 33.4 ± 1.5 HU) compared with EID-CT (GM: 45.1 ± 1.6 HU; WM: 37.4 ± 1.6 HU; P < .001). Standard deviations of HU were also lower in PCD-CT (GM and WM both P < .001) and CNR was significantly higher in PCD-CT compared with EID-CT (P < .001). GWRs were not significantly different across both modalities (P > .99). In an age-matched subset (n=157 patients from both cohorts), all findings were replicated.
Conclusions:
This comprehensive comparison of HU in cerebral GM and WM revealed substantially reduced image noise and an average offset with lower HU in PCD-CT while the ratio between GM and WM remained constant. The potential need to adapt windowing presets based on this finding should be investigated in future studies.
More Related Videos
15:10A Dual Tracer PET-MRI Protocol for the Quantitative Measure of Regional Brain Energy Substrates Uptake in the Rat
Published on: December 28, 2013
09:21Human Brown Adipose Tissue Depots Automatically Segmented by Positron Emission Tomography/Computed Tomography and Registered Magnetic Resonance Images
Published on: February 18, 2015
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies III: Computed Tomography