Related Experiment Video
Updated: Sep 17, 2025

Fully Endoscopic Mitral Valve Repair with Percutaneous Cannulation of Groin Vessels
Published on: May 26, 2023
VALVE-IN-VALVE IMPLANTATION GONE SOUTH: A CASE REPORT
Robert Bernat1,2, Volker Windmüller1, Vasileios Leivaditis1
1Westpfalz-Klinikum Kaiserslautern, Germany.
None:
Valve-in-Valve (ViV) procedures for failed surgical bioprostheses are becoming increasingly common. We present a case of a 73-year-old male patient with significant comorbidities who underwent a primary transcatheter aortic valve implantation (TAVI) with an Evolut 29 mm (Medtronic, Minneapolis, MN, USA) valve. This TAVI-prosthesis implantation failed due to the inability to fix the valve at the appropriate height to achieve an optimal hemodynamic result. The valve had to be snared and pulled into an adequate position and was then fixed by an implantation of another TAVI-prosthesis (29 mm Sapien S3, Edwards Lifesciences, Irvine, CA, USA) with optimal final result. The patient was discharged after significant clinical improvement and without further complications.
Related Concept Videos
Mitral Valve Prolapse II: Assessment and Management
Mitral Valve Prolapse I: Introduction
Mitral Stenosis I: Introduction
Mitral Regurgitation I: Introduction
Mitral Valve Prolapse III: Nursing Management
Mitral Regurgitation III: Medical Management