Related Experiment Video
Updated: Sep 17, 2025

Tracking the Mammary Architectural Features and Detecting Breast Cancer with Magnetic Resonance Diffusion Tensor Imaging
Published on: December 15, 2014
Development of a carrier device for dental-dedicated magnetic resonance imaging
Matheus Sampaio-Oliveira1,2, Matheus L Oliveira1, Rubens Spin-Neto2
1Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, University of Campinas, Piracicaba, SP, Brazil.
Purpose:
The aim of this study was to develop and evaluate a carrier device for dental-dedicated magnetic resonance imaging (ddMRI).
Materials And Methods:
The carrier device comprised 5 glass test tubes, which were vertically positioned within a glass beaker and filled with air, distilled water, 1.5% agar, nickel nitrate [Ni(NO3)2] in 1.5% agar, or 1000 g·L-1 dipotassium phosphate (K2HPO4). The beaker was filled with distilled water, a 0.3 g·L-1 Ni(NO3)2 aqueous solution, or a 1000 g·L-1 K2HPO4 aqueous solution. The device was scanned using a proton density turbo-spin-echo pulse sequence on a ddMRI system equipped with a dental-dedicated radiofrequency surface coil. Triplicate scans were performed for each combination of tube fillings and beaker solutions, yielding a total of 45 image volumes. Quantitative image metrics were then assessed.
Results:
The developed carrier device, composed of carrier vials filled with 1.5% agar surrounded by a 1000 g·L-1 K2HPO4 aqueous solution, was identified as the best option for ddMRI quality assessments.
Conclusion:
The proposed carrier device represents a promising method for embedding dental materials and other specimens, thereby facilitating the evaluation of their behaviour in ddMRI.