Temperature and Humidity During Different Neonatal Respiratory Support Modes: An In-Vitro Performance of Four Heated
Emanuela Zannin1, Camilla Rigotti1, Stefano Robbiani2
1Neonatal Intensive Care Unit, Fondazione IRCCS San Gerardo dei Tintori, Monza, Italy.
Background:
Adequate heating and humidification of inspired gas are crucial in preterm infants requiring respiratory support. Our objective was to compare the temperature (T) and absolute humidity (AH) achieved with different neonatal heated humidifiers commonly used during different respiratory support modes and respiratory settings.
Methods:
Four neonatal heated humidifiers were tested at different gas flow rates and during conventional and high-frequency oscillatory ventilation at clinically relevant settings. The patient was simulated using a neonatal test lung placed in a neonatal incubator warmed at 37°C. Temperature and relative humidity were measured using a thermohygrometer placed proximal to the test lung and used to calculate AH.
Results:
Measured temperature and humidity varied among heated humidifiers and respiratory support settings. The bias flow was the primary determinant of the measured temperature and humidity: temperature and AH significantly increased with flow. When the flow through the chamber was below 5 L/min the MR950 produced T < 36.7°C and AH < 37 g/m3 proximal to the patient. The Hydraltis 9500neo with chamber and proximal temperature set to 38°C and 40°C and the MR850 produced AH > 44 g/m3 in most conditions.
Conclusion:
The temperature and humidity proximal to the patients' airways might vary considerably based on the heated humidifiers and the flow through the chamber. Further clinical studies are needed to evaluate the effect of such variability on the patients.
More Related Videos
05:45Delivery of In Vivo Acute Intermittent Hypoxia in Neonatal Rodents to Prime Subventricular Zone-derived Neural Progenitor Cell Cultures
Published on: November 2, 2015
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
Related Concept Videos
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Acute Respiratory Failure-IV
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...