Hand dynamometry: Does the device used for measurement matter?
Diana García Sastre1, Begoña Pintor de la Maza1, María López Melgar1
1Endocrinology and Nutrition, Complejo Asistencial Universitario de León, León, Spain.
Introduction:
Handgrip dynamometers measure muscle strength through the grip strength of the dominant hand. The purpose of this study is to compare the Dynx, Saehan, and Smedlay dynamometers with the Jamar dynamometer (gold standard) by evaluating the correlation and concordance of their measurements to determine if it is indifferent to use different dynamometers.
Material And Methods:
Cross-sectional study conducted between 2020 and 2021. We evaluated a group of patients (GP) from the Endocrinology and Nutrition consultations and a control group (CG) of health personnel. Dominant hand grip strength was measured with four different dynamometers Jamar, Dynx, Saehan and Smedlay, in random order. Statistical analysis was performed with SPSS using Student's t-test, intraclass correlation coefficient (ICC), and Bland-Altman plots.
Results:
A total of 124 subjects were included; 91 GP (60.4% women), mean age 54.9 (SD14.2) years. ICC showed the best correlation for Saehan in GP (0.977 p < 0.001) and in GC (0.960 p < 0.001). Dynx correlated better in GP than in GC (0.967 vs. 0.900, p < 0.001) and Smedlay had the worst correlation in GC (0.812 p < 0.001). The dynamometer with the lowest bias (-0.39, SD 3.11) and better limits of agreement (LOA) in both groups (GP -0.39, GC -0.51) was the Saehan CONCLUSION: Despite good measurement correlation, there is bias and insufficient agreement between different dynamometers, so they should not be used interchangeably.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)