Related Concept Videos

Equipments Used To Measure Blood Pressure01:30

Equipments Used To Measure Blood Pressure

3.4K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
3.4K
Assessment of blood pressure in brachial artery(one-step method)01:15

Assessment of blood pressure in brachial artery(one-step method)

1.4K
This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
1.4K
Assessment of blood pressure in brachial artery(two-step method)01:23

Assessment of blood pressure in brachial artery(two-step method)

2.1K
Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
2.1K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Long-Term Body Composition Trajectories After Bariatric Surgery: A 5-Year Comparative Study of Biliopancreatic Diversion, Roux-en-Y Gastric Bypass, and Sleeve Gastrectomy.

Journal of clinical medicine·2026
Same author

Prevalence of malnutrition, sarcopenia, and sarcopenic obesity in patients with heart failure.

Endocrinologia, diabetes y nutricion·2026
Same author

Associations Between Klotho/FGF-Related Protein Expression in Peripheral Blood Mononuclear Cells, Inflammation, and Muscle Function in Middle-Aged Adults with Obesity: A Pilot Study.

International journal of molecular sciences·2026
Same author

Pembrolizumab-induced dual endocrine disruption: A case report.

Endocrinologia, diabetes y nutricion·2025
Same author

Absolute and adjusted hand grip strength values in obese patients.

Endocrinologia, diabetes y nutricion·2025
Same author

Influence of Klotho Protein Levels in Obesity and Sarcopenia: A Systematic Review.

International journal of molecular sciences·2025

Related Experiment Video

Updated: May 6, 2026

Measurement of Spatial Stability in Precision Grip
09:36

Measurement of Spatial Stability in Precision Grip

Published on: June 4, 2020

3.3K

Hand dynamometry: Does the device used for measurement matter?

Diana García Sastre1, Begoña Pintor de la Maza1, María López Melgar1

  • 1Endocrinology and Nutrition, Complejo Asistencial Universitario de León, León, Spain.

Endocrinologia, Diabetes Y Nutricion
|July 10, 2025
PubMed
Summary

Different handgrip dynamometers show good measurement correlation but significant bias. Saehan, Dynx, and Smedlay dynamometers should not be used interchangeably with the Jamar dynamometer due to insufficient agreement.

Keywords:
Bland AltmanCorrelación intraclaseDinamómetroDynamometerFuerza de agarre (HGS)Hand Grip Strength (HGS)Intraclass correlationSarcopenia

More Related Videos

Measuring the Motor Aspect of Cancer-Related Fatigue using a Handheld Dynamometer
07:22

Measuring the Motor Aspect of Cancer-Related Fatigue using a Handheld Dynamometer

Published on: February 20, 2020

5.9K
Measurement of the Hand Transmitted Vibration of the Human Hand Arm System During Operation of a Hand Tractor
09:35

Measurement of the Hand Transmitted Vibration of the Human Hand Arm System During Operation of a Hand Tractor

Published on: June 16, 2021

4.5K

Related Experiment Videos

Last Updated: May 6, 2026

Measurement of Spatial Stability in Precision Grip
09:36

Measurement of Spatial Stability in Precision Grip

Published on: June 4, 2020

3.3K
Measuring the Motor Aspect of Cancer-Related Fatigue using a Handheld Dynamometer
07:22

Measuring the Motor Aspect of Cancer-Related Fatigue using a Handheld Dynamometer

Published on: February 20, 2020

5.9K
Measurement of the Hand Transmitted Vibration of the Human Hand Arm System During Operation of a Hand Tractor
09:35

Measurement of the Hand Transmitted Vibration of the Human Hand Arm System During Operation of a Hand Tractor

Published on: June 16, 2021

4.5K

Area of Science:

  • Biomedical Engineering
  • Clinical Measurement Science
  • Rehabilitation Medicine

Background:

  • Handgrip dynamometers are crucial for assessing muscle strength.
  • The Jamar dynamometer is considered the gold standard for grip strength measurement.
  • Evaluating alternative dynamometers is essential for clinical practice and research.

Purpose of the Study:

  • To compare the measurement consistency of Dynx, Saehan, and Smedlay dynamometers against the Jamar dynamometer.
  • To assess the correlation and concordance between different handgrip strength measurement devices.
  • To determine if interchangeable use of these dynamometers is appropriate.

Main Methods:

  • A cross-sectional study involving 124 participants (91 patients, 33 controls) between 2020-2021.
  • Grip strength was measured using Jamar, Dynx, Saehan, and Smedlay dynamometers in a randomized order.
  • Statistical analysis included intraclass correlation coefficient (ICC) and Bland-Altman plots.

Main Results:

  • Saehan demonstrated the highest correlation in both patient (0.977) and control groups (0.960).
  • Dynx showed strong correlation (0.967 in patients, 0.900 in controls).
  • Saehan exhibited the lowest bias (-0.39) and best limits of agreement, but significant bias and insufficient agreement were found between devices.

Conclusions:

  • While high correlation exists, significant bias and lack of agreement necessitate caution.
  • The tested dynamometers (Saehan, Dynx, Smedlay) are not interchangeable with the Jamar dynamometer.
  • Clinical decisions based on grip strength measurements should consider the specific dynamometer used.