Dosimetric evaluation of a novel automated noncoplanar volumetric modulated arc therapy technique for treating optic nerve sheath meningiomas
Related Experiment Video
Updated: May 1, 2026

Role of Diffusion MRI Tractography in Endoscopic Endonasal Skull Base Surgery
Published on: July 5, 2021
Zhenyu Xiong1,2,3, Chingyun Cheng4, Lili Zhou3
1Department of Radiation Oncology, Rutgers Cancer Institute of New Jersey, Rutgers, The State University of New Jersey, New Brunswick, NJ, United States.
Purpose:
This study aimed to evaluate the dosimetric outcomes for the target and organs at risk (OARs) in patients with optic nerve sheath meningiomas (ONSMs), comparing HyperArc (HA), a novel automated noncoplanar volumetric modulated arc therapy (VMAT) technique, with two other advanced VMAT techniques.
Methods:
Nine patients with ONSMs were re-planned using three radiotherapy techniques: HA employing four preconfigured noncoplanar partial arcs on the Varian TrueBeam, a two-arc coplanar VMAT on the Varian TrueBeam (TB-VMAT), and a two-arc coplanar VMAT on the Varian Halcyon (HAL-VMAT). All treatment plans aimed to deliver 50.4 Gy in 28 fractions to the planning target volume (PTV) while minimizing dose to OARs. The planning process began by applying identical preset optimization templates for each plan, followed by iterative refinements of objectives and priorities to accommodate individual plan requirements. All plans were normalized to ensure that 100% of the prescription dose covered 95% of the PTV. Dosimetric evaluation included PTV metrics (D98%, Dmean, Dmax, and Dmin), the Paddick Conformity Index (PCI), the International Commission on Radiation Units and Measurements Report 83 (ICRU-83) homogeneity index (HI), the gradient index (GI), and doses to OARs for each technique. Statistical significance was assessed using the Wilcoxon signed-rank test with a p-value threshold of < 0.05.
Results:
HA plans demonstrated superior dosimetric indices for PTV, as indicated by the highest D98% (50.24 ± 0.05 Gy) and the lowest Dmax (53.20 ± 0.23 Gy), HI (0.04 ± 0.00), and GI (3.56 ± 0.58) values (p < 0.05). These results indicated superior target coverage and a more homogeneous dose distribution. Furthermore, HA plans achieved the lowest maximum dose values for the following OARs: lenses, hippocampi, contralateral optic nerve, and contralateral retina (p < 0.05), thereby optimally sparing these critical structures. No significant differences were observed across techniques regarding Dmean, Dmin, PCI, or maximum dose to the ipsilateral optic nerve, ipsilateral retina, and optic chiasm.
Conclusions:
HA plans demonstrated superior dosimetric performance, ensuring adequate target coverage, reduced PTV hotspots, and better OAR protection compared to coplanar VMAT plans on the Varian TrueBeam and Halcyon. These advantages suggest that.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
08:57Quantification of Optic Nerve Cross Sectional Area on MRI: A Novel Protocol using Fiji Software
Published on: September 4, 2021
Related Concept Videos
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment