Pulse Oximetry Discrepancies and Occult Hypoxemia in ICU Patients: Predictors and Clinical Outcomes
Saikou Saidy1, Ali Iqbal1, Saqib H Baig1
1Division of Pulmonary, Allergy and Critical Care, Jane and Leonard Korman Respiratory Institute, Sidney Kimmel Medical College, Thomas Jefferson University, Philadelphia, PA, USA.
Abstract:
BackgroundPulse oximeters sometimes fail to accurately reflect arterial oxygen saturation (SaO2), particularly in darker-skinned patients resulting in undiagnosed hypoxemia, potentially delaying recognition and appropriate interventions.Research QuestionWe aimed to evaluate the prevalence and predictors of SpO2-SaO2 discrepancies, particularly occult hypoxemia, and to assess their association with clinical outcomes in ICU patients.Study Design and MethodsWe conducted a retrospective cohort analysis using the Blood-gas and Oximetry Linked Dataset (BOLD), analyzing critically ill patients from the eICU-CRD database (2014-2015). Patients with paired SpO2-SaO2 measurements within five minutes were included. We identified SpO2-SaO2 discrepancies as a difference of >2.99% and defined occult hypoxemia as an arterial partial pressure of oxygen (PaO2) < 60 mm Hg or SaO2 < 89% with an SpO2 > 88%. The primary outcomes included ICU length of stay (LOS), Sequential Organ Failure Assessment (SOFA) score, and in-hospital mortality.ResultsAmong 36,280 ICU patients, 23.6% had SpO2-SaO2 discrepancies, and 4.7% had occult hypoxemia. Black patients were overrepresented in both groups, with an adjusted odds ratio (aOR) of 1.35 (95% CI: 1.25-1.47) for discrepancy and 1.22 (95% CI: 1.04-1.47) for occult hypoxemia. Higher BMI, lower pH, elevated creatinine, and higher Charlson Comorbidity Index scores were also significant predictors. Patients with discrepancies had worse clinical outcomes, including increased SOFA scores in the following 24 h (β = 0.31; p < .0001) and higher in-hospital mortality (aOR 1.15; p < .0001). Occult hypoxemia was associated with even worse outcomes, including a longer ICU LOS (IRR 1.12; p < .0001) and significantly increased mortality (aOR 1.73; p < .0001).InterpretationOne in four critically ill patient in our cohort experienced SpO2-SaO2 discrepancy which is associated with adverse clinical outcomes. Black race, obesity, and higher comorbidity burden were significant predictors of these discrepancies. Our findings emphasize the need for more rigorous clinician oversight in the use of this technology.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Hypoxia
Types of Hypoxia
There are four primary types of hypoxia, each resulting from a different cause:
1. Anemic hypoxia: This type occurs due to insufficient oxygen delivery caused by a lack of red blood cells (RBCs) or RBCs with abnormal or...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...