Related Experiment Video
Updated: Sep 15, 2025

CO2-Lasertonsillotomy Under Local Anesthesia in Adults
Published on: November 6, 2019
Redefining Safe Practices in Airway TruBlue Laser Surgery-A Systematic Evaluation
Ariel Roitman1,2, Marcus Wilson3, Andrew J Bowen1
1Department of Surgery, Division of Otolaryngology-Head and Neck Surgery, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.
Objectives:
Fire risk during laser airway surgery is well-documented, but there is limited information regarding this risk with the novel TruBlue laser. This study evaluates its safety under different experimental conditions.
Methods:
Using a cadaveric porcine larynx and lungs model, we conducted trials with laser-safe endotracheal tubes (LSETTs), high-flow ventilation (HFV), and jet ventilation. We varied distance, FiO2 levels, and laser power settings to observe the occurrence of uncontrolled fires, brief flames, and sparks.
Results:
LSETT trials showed no fire events. Pledget fires were the only uncontrolled fires in HFV, at 100%, 70%, and 50% FiO2. Brief flames occurred only with 4 W continuous settings at 100% FiO2. Innocuous spark ignition rates were significantly increased by increased wattage (p < 0.05). In jet ventilation trials, brief flames appeared across a wider range of FiO2 levels and settings. HFV and jet ventilation groups showed a greater frequency and earlier occurrence of sparks and brief flames, in contrast to the event-free LSETT group (p < 0.05).
Conclusion:
Using the TruBlue laser with a fully inflated LSETT cuff positioned at least 0.3 cm below the operating area while delivering 100% oxygen was consistently safe in our trials. Uncontrolled fires occurred only when targeting a pledget under HFV, confirming extrinsic fuel as the main risk and highlighting the safety of the TruBlue laser with HFV and jet ventilation in its absence. Compared to HFV, jet ventilation trials showed a higher frequency of brief flames and sparks. A 60-s continuous application of the laser at 100% FiO2 is safe.
Level Of Evidence:
NA.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...
Tracheostomy Suctioning II: Procedure
Cardiopulmonary Resuscitation II: ACLS Airway Management
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes:
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask...