The QuickSee autorefractor reliably measures the accommodative response in children compared with the Shin-Nippon
Jane M Fulton1, Rebecca E Leighton1, Tuyishime Didier Fidele Uwacu1
1Centre for Optometry and Vision Science, School of Biomedical Sciences, Ulster University, Coleraine, UK.
Summary
The QuickSee autorefractor shows promise for measuring accommodative response in children without significant refractive errors. Its measurements closely align with the gold-standard Shin-Nippon device, suggesting potential for vision screening applications.
Area of Science:
- Ophthalmology
- Pediatric Optometry
- Vision Science
Background:
- Accurate measurement of accommodative response is crucial for understanding visual development and function in children.
- The QuickSee autorefractor is a novel device for assessing refractive status and accommodative function.
- Comparison with a gold-standard device is necessary to validate new autorefractor performance.
Purpose of the Study:
- To evaluate the QuickSee autorefractor's ability to measure accommodative response in children.
- To compare QuickSee measurements with those obtained from the Shin-Nippon autorefractor, a recognized gold-standard.
- To assess the inter-instrument agreement and reliability of the QuickSee for accommodative response assessment.
Main Methods:
- Recruited children aged 5-7 years without significant ametropia.
- Measured refractive status using QuickSee and Shin-Nippon autorefractors at distance and near demands (2, 3, and 4 D).
- Calculated accommodative response and analyzed inter-instrument agreement using Bland-Altman methods and repeated measures ANOVA.
Main Results:
- Forty-nine children participated, with a median spherical equivalent refraction of +0.38 D.
- Accommodative responses measured by QuickSee and Shin-Nippon showed close agreement across different accommodative demands.
- No significant proportional bias or significant differences in accommodative response measurements were found between the two devices (p ≥ 0.19).
Conclusions:
- The QuickSee autorefractor demonstrates good agreement with the Shin-Nippon device for measuring accommodative response in children with normal refractive status.
- The findings suggest the QuickSee holds potential for use in pediatric vision screening and assessment of accommodative function.
- Further research is warranted to explore its utility across a broader spectrum of refractive errors and in diverse clinical settings.