An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
The Utility of Evaluating Primary Stability of Implants Using Accumulated Insertion Torque Values: a Preliminary
Reiko Kobatake1, Kazuya Doi1, Tomoko Izumikawa1
1Department of Advanced Prosthodontics, Hiroshima University Graduate School of Biomedical and Health Sciences, HiroshimaJapan.
This study introduces accumulated torque value (ATV) as a new method for evaluating primary implant stability. ATV correlates with peak torque value and implant stability quotient, showing promise for dental implant assessment.
Area of Science:
- Dental Implantology
- Biomaterials Engineering
- Surgical Instrumentation
Background:
- Primary implant stability is crucial for successful osseointegration.
- Current methods for assessing primary stability have limitations.
- Novel, reliable methods are needed to evaluate implant stability accurately.
Purpose of the Study:
- To investigate the potential of accumulated torque value (ATV) as a novel metric for assessing primary dental implant stability.
- To compare ATV with peak torque value (PTV) and implant stability quotient (ISQ).
Main Methods:
- Utilized simulated bone blocks of varying densities (D1-D4) and two implant sizes (42-10ST, 37-6ST).
- Measured implant insertion torque at 0.05-second intervals, recording peak torque value (PTV) and accumulated torque value (ATV).
- Assessed implant stability quotient (ISQ) post-insertion and analyzed correlations between ATV, PTV, and ISQ.
Main Results:
- ATV demonstrated a positive correlation with simulated bone density.
- Strong correlations were found between ATV and PTV (r=0.99 for both implant sizes).
- Significant correlations were also observed between ATV and ISQ (r=0.81-0.83).
Conclusions:
- Accumulated torque value (ATV) is influenced by bone density, reflecting primary implant stability.
- ATV shows a strong correlation with peak torque value and existing implant stability measures.
- ATV presents a promising, novel method for evaluating primary dental implant stability.
More Related Videos
07:32Author Spotlight: 3D Movement Assessment of Maxillary Posterior Teeth in Clear Aligner Treatment
Published on: February 23, 2024
09:07An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
Related Concept Videos
Residual Stresses in Circular Shafts
Stability of structures
Screw: Problem Solving
To evaluate the...
Stress Concentrations in Circular Shafts