We present our experience treating ocular tumors in a standard pencil beam scanning (PBS) gantry room without apertures, which could broaden access to proton therapy for patients with ocular cancer globally. Besides, this study explores the dosimetric benefits of beam-specific apertures.
Methods And Materials:
We retrospectively evaluated 11 consecutive patients with uveal melanoma treated in a clinic gantry room. The dose deviations between the planned and received by the patient were investigated by assessing the forward calculation of the treatment plan on the synthetic computed tomography of cone beam computed tomography. Each plan was forward calculated with a beam-specific brass aperture (BSA) using a Monte Carlo algorithm to explore dosimetric improvements. We compared the plan quality to the delivered plan (DP) using target coverage (D95%) and mean/maximum doses to the adjacent organs.
Results:
A close agreement between the planned and delivered dose was achieved, with D95% deviations within 3.6% for all treatments, maintaining dose constraints for critical organs. Similar target coverage was reached, with D95% at 101% ± 1.0% (DP) and 101% ± 3.2% (BSA). BSA was effective (P < .05) in reducing the mean [D Mean (DP, BSA)Gy] and maximum [D Max (DP, BSA)Gy] dose to organs: retina D Mean (37.7, 29.5), cornea D Mean (10.7, 2.4), conjunctiva D Mean (13.6, 4.1), lacrimal gland D Mean (25.5, 14.1), optic nerve D Mean (19.6, 13.1), lens D Max (22.4, 8.5), cornea D Max (24.4, 10.2), eyebrow D Max (15.3, 6.8). BSA lowered the mean dose to surrounding organs and significantly decreased the maximum dose to nonabutting organs (lens, cornea, eyebrow), but had little impact on the maximum dose to the abutting organs (retina, optic nerve).
Conclusions:
We demonstrate the successful implementation of ocular proton treatment with a standard PBS gantry beamline without apertures. The beam-specific apertures effectively reduced doses to the organs adjacent to the target in the PBS proton treatment while maintaining similar target coverage. This approach offers an opportunity to expand access to ocular proton therapy widely.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment