The 10 m collimated Rowland spectrometer at the MAX IV Veritas beamline
Marcus Agåker1, Carl Johan Englund2, Victor Ekholm1
1MAX IV Laboratory, Lund University, PO Box 118, SE-22100 Lund, Sweden.
None:
A grazing-incidence Rowland spectrometer with a collimating element for resonant inelastic soft X-ray scattering at the Veritas beamline at MAX IV, Lund, Sweden, is presented. The instrument is designed to operate in the 250-950 eV range, providing >30000 resolving power in first order. The inherent temporal resolution of the delay-line detector enables time-resolved experiments on the 270 ps time-scale and also allows for so-called time-gating which significantly improves the signal-to-noise ratio. The instrument is able to rotate around the sample position in the horizontal plane, which enables measurements of the momentum transfer in the X-ray scattering event. The instrument is prepared for installation of a polarimeter to provide full characterization of energy, momentum and polarization of the scattered photons. The design is evaluated using spectral measurements, ray-tracing, vibration and motion measurements.
Related Concept Videos
UV–Vis Spectrometers
IR Spectrometers
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Emission Spectroscopy: Instrumentation
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....