Characteristics of a Miniature Fiber-Optic Inorganic Scintillator Detector for Electron-Beam Therapy Dosimetry
Zhuang Qin1, Ziyin Chen2,3, Bo He3
1School of Electronic Engineering, Heilongjiang University, Harbin 150080, China.
Abstract:
Over the past few decades, electron beams have been widely used to treat malignant and benign tumors located in the superficial regions of patients. This study utilized an inorganic scintillator (Gd2O2S:Tb)-based radiation detector to test its response characteristics in an electron-beam radiotherapy environment, in order to determine the application potential of this detector in electron-beam therapy. Owing to the extremely high time resolution of this inorganic scintillator detector (ISD), it is even capable of measuring the pulse information of electron beams generated by the accelerator. The results indicate that for certain accelerator models, such as the IX3937, the pulse pattern of the output electron beam is notably different from that during the output of X-rays, showing no significant periodicity. The experimental results also demonstrate that this ISD exhibits excellent repeatability and dose linearity (R2 of 0.9993) when measuring electron beams. Finally, the PDD (Percentage Depth Dose) curves and OAR (Off-Axis Ratio) curves of the ISD were also tested under electron-beam conditions at 6 MeV and 9 MeV, respectively.


