Assessing the Quantitative Performance of Atmospheric Solids Analysis Probe-Mass Spectrometry
Alisha Henderson1, Stephanie Rankin-Turner2, James C Reynolds3
1School of Sport, Exercise and Health Sciences, Loughborough University, Loughborough, UK.
Rationale:
Atmospheric solids analysis probe-mass spectrometry (ASAP-MS) is an established ambient ionisation technique that allows for the direct and rapid analysis of samples without chromatographic separation. Consequently, applications that typically benefit from ambient ionisation approaches can achieve improved sample throughput and thus improved potential for in situ testing. Previous reports have contrasted in viewpoints on the ability for ASAP-MS to provide reliable quantitative data. Critically, in-depth data exploring the quantitative capabilities of ASAP-MS are currently lacking.
Methods:
Here, a series of experiments were performed to assess the quantitative performance of ASAP-MS using a proof-of-concept single analyte (caffeine) approach. Analytical precision, accuracy, linearity and sensitivity were investigated using numerous variables, including sample deposition method (i.e., directly placing the probe into the sample vs. pipetting the sample onto the probe) and deposition volume, as well as the presence of a series of different internal standard approaches.
Results:
The data acquired demonstrated that the use of a positive displacement pipette and an isotopically labelled (structure-matched) internal standard provided an optimal approach for quantitative reliability, albeit at levels often below the standards set for traditional chromatography-based quantitative assays. The investigations were performed across a concentration range of 50-5000 ng/mL. Whereas measurable responses were seen across the full range in most approaches, limitations in sensitivity were identified and reduced quantitative performance statistics were noted at concentrations below 1000 ng/mL.
Conclusions:
These experiments demonstrate, for the first time, a comprehensive investigation into the quantitative performance of ASAP-MS using caffeine as the example analyte. These data offer insight into the current strengths and limitations of quantitative analyses using ASAP-MS and aim to provide practical recommendations to optimise quantitative approaches using this technique.
More Related Videos
08:43PTR-ToF-MS Coupled with an Automated Sampling System and Tailored Data Analysis for Food Studies: Bioprocess Monitoring, Screening and Nose-space Analysis
Published on: May 11, 2017
05:47Sample Preparation for Probe Electrospray Ionization Mass Spectrometry
Published on: February 19, 2020
Related Concept Videos
Mass Spectrometers
Mass Analyzers: Overview
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Tandem Mass Spectrometry
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...
Mass Spectrometry: Overview