Related Experiment Video
Updated: Sep 13, 2025

Improving IV Insulin Administration in a Community Hospital
Published on: June 11, 2012
Switched Controllers in Fully Closed Loop Insulin Delivery Systems: Reducing the Trade-Off Between Prandial Control
Fernando Leonel Da Rosa Jurao1, Emilia Fushimi1, Fabricio Garelli1
1Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales - LEICI (UNLP-CONICET), Facultad de Ingeniería, Universidad Nacional de La Plata, La Plata, Argentina.
Background:
One of the main challenges in control algorithm design for full closed-loop automated insulin delivery systems is the trade-off between the effective compensation of meal-related disturbances and ensuring user safety during the postprandial and fasting periods.
Methods:
This paper proposes and evaluates the performance of a switched tuning strategy, a promising but relatively underexplored solution in this domain. This method employs two distinct tunings of a primary control algorithm: an aggressive tuning for meal compensation and a conservative tuning for fasting periods. The analysis considers implementing the switched strategy for three control algorithms: model predictive control and proportional-derivative control, both widely used for glucose regulation, and a linear quadratic Gaussian control, an optimal algorithm previously validated in clinical settings under a switched structure. Additionally, to obtain a more comprehensive understanding of the switched strategy implications, two nonswitched controllers are implemented for each control algorithm: an aggressive and a conservative tuning strategy.
Results:
The switched strategy significantly improves the trade-off between meal compensation and safety, increasing the time within the target range of 70-180 [mg/dL] for all three algorithms. For proportional-derivative control, the time in range increases from 69.1% with the conservative tuning and 83.1% with the aggressive to 86.6% with the switched structure. For model predictive control, the improvement is from 73.4% and 74.1% to 85.8%. Last, linear quadratic Gaussian control increases from 65.0% and 70.4% to 85.6%.
Conclusion:
The findings suggest that the switched strategy may be a feasible and straightforward approach for enhancing meal compensation without increasing the risk of postprandial hypoglycemia in people with diabetes.
More Related Videos
09:08Three-dimensional Printing of Thermoplastic Materials to Create Automated Syringe Pumps with Feedback Control for Microfluidic Applications
Published on: August 30, 2018
08:32Studying the Hypothalamic Insulin Signal to Peripheral Glucose Intolerance with a Continuous Drug Infusion System into the Mouse Brain
Published on: January 4, 2018
Related Concept Videos
Open and closed-loop control systems
An open-loop control system operates without feedback from the output. It consists of two primary elements: the controller and the controlled process. The controller receives an input signal...
PI Controller: Design
PID Controller
Controller Configurations
Control-system compensation involves various configurations, most commonly series or cascade compensation, in which the controller...
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...