Towards Seamless Integration of Magnetic Tracking Into Fluoroscopy-Guided Interventions
IEEE Transactions on Bio-Medical Engineering
|August 7, 2025
Summary
This study introduces a radiolucent magnetic field generator for improved magnetic tracking in X-ray guided procedures, enhancing depth perception and reducing radiation exposure during interventions.
Area of Science:
- Medical Imaging and Interventional Radiology
- Biomedical Engineering
- Surgical Navigation Technologies
Background:
- Fluoroscopy-guided interventions suffer from limited depth perception and high radiation exposure due to 2D imaging.
- Integrating magnetic tracking (MT) offers potential solutions but faces integration and accuracy challenges.
- Current research and practice have under-explored seamless MT integration into fluoroscopy workflows.
Purpose of the Study:
- To develop and evaluate a radiolucent magnetic field generator (FG) prototype for seamless magnetic tracking integration in fluoroscopy-guided interventions.
- To address technical challenges in C-arm pose estimation, fluoro-CT registration, and 3D navigation for enhanced interventional accuracy.
- To demonstrate the clinical applicability and efficacy of MT-assisted navigation in a representative endoleak repair phantom study.
Main Methods:
- Designed a two-layer radiolucent FG mounting frame for C-arm X-ray system compatibility.
- Incorporated external aluminum fiducials to facilitate accurate C-arm pose estimation and robust fluoro-CT registration.
- Evaluated MT accuracy with fiducials and C-arm presence, assessed fluoro-CT registration accuracy, and performed a phantom insertion experiment for endoleak repair.
Main Results:
- Experimental evaluation confirmed no clinically significant impact of aluminum fiducials or the C-arm on MT accuracy.
- Achieved high accuracy (mean projection distance ≈ 0.7 mm), robustness, and generalizability in fluoro-CT registration.
- Demonstrated needle insertion accuracy between 2-3 mm in a phantom study using real-time 2D and 3D navigation.
Conclusions:
- The developed radiolucent FG and integrated MT approach are effective and clinically applicable for fluoroscopy-guided interventions.
- This study represents the first integration of a radiolucent FG into a fluoroscopy-guided workflow, paving the way for improved interventional procedures.
- The proposed methods overcome key technical challenges, enhancing navigation accuracy and potentially reducing radiation exposure in interventional radiology.