Hemolysis Induced by Pulsed-Field Ablation of Atrial Arrhythmias: A Comparative Analysis of Current Systems
Johannes Bruss1, Thomas Kueffer1, Hildegard Tanner1
1Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.
Introduction:
Pulsed-field ablation (PFA) is an emerging technology associated with dose-dependent hemolysis as a recently recognized side effect. This study aimed to compare hemolysis levels and assess dose-dependency across three PFA systems: a pentaspline catheter (PSC), a lattice-tip focal catheter (LTFC), and a variable loop circular catheter (VLCC).
Methods:
Patients treated for atrial arrhythmias with the LTFC (n = 29) or the VLCC (n = 30) were included from a prospective registry. A matched cohort of patients treated with the PSC (n = 28) was recruited from the same registry. Creatinine levels and markers of hemolysis were measured pre-ablation and 1 day postablation.
Results:
Haptoglobin levels decreased significantly more with the PSC and VLCC compared to the LTFC (-0.65 [-0.76, -0.49] g/L; -0.56 [-0.78, -0.43] g/L, -0.21 [-0.32, -0.1] g/L, respectively; p < 0.001 for both). Per-application decreases in haptoglobin also differed (-17.5 [-20.38, -13.58] mg/L, -24.35 [-36.36, -17.92] mg/L, -3.61 [-5.98, -2.13] mg/L, respectively; p < 0.001 for both). There was no significant difference in haptoglobin decrease between the PSC and VLCC per procedure (p = 1.0). Haptoglobin decrease per application was significantly larger with the VLCC compared to the PSC (p = 0.0048). Per procedure LDH increase followed a similar trend (49 [18, 81.25] U/L; 14 [6, 60] U/L; 13 [-4, 46] U/L; respectively; p = 0.037). No hemolysis-related complications were observed.
Conclusions:
Hemolysis levels vary significantly among PFA platforms. Focal PFA catheters induce less hemolysis per procedure and application compared to large-footprint catheters.
Related Concept Videos
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Heart Failure II: Pathophysiology