Dental photon-counting computed tomography for the assessment of Peri-Implant structures
Maurice Ruetters1, Christian Mertens2, Holger Gehrig3
1Department of Operative Dentistry, Heidelberg University, University Hospital Heidelberg, Im Neuenheimer Feld 400, 69120, Heidelberg, Germany. Maurice.ruetters@med.uni-heidelberg.de.
Purpose:
To assess the diagnostic performance of photon-counting computed tomography (PCCT) in the imaging of peri-implant bone structures and to compare it quantitatively and qualitatively to cone-beam computed tomography (CBCT).
Methods:
Thirty titanium implants were placed in ten porcine mandibles. CBCT and PCCT scans were acquired and compared quantitatively regarding image noise and CT-values. Additionally bone thickness was compared to a gold standard at 60 standardized locations by one calibrated investigator in both modalities. Measurement accuracy was assessed by Bland-Altman analysis. Two experienced raters performed qualitative assessments of anatomic structures around the implant using a 5-point visibility scale. These included the bone-implant interface around the implant surface, the bone at the implant shoulder as well as the oral and vestibular bone lamella. Inter-rater agreement was assessed using ICC.
Results:
Across all evaluated implants, CT-values in a soft-tissue region of interest adjacent to the implant increased by 11.7 ± 3.9% for CBCT acquisitions, whereas they decreased by 5.3 ± 1.3% for PCCT acquisitions. Similarly, image noise in the respective ROIs is increased by a factor of 63 ± 13% in case of CBCT acquisitions and only by 23 ± 5% in case of PCCT acquisitions. Bone thickness deviations were smaller for PCCT (mean ± SD: 0.06 ± 0.08 mm) than for CBCT (0.39 ± 0.34 mm). Qualitative assessments consistently favored PCCT (p < 0.05) with excellent inter-rater reliability (ICC > 0.75 ) in almost all categories.
Conclusions:
PCCT enables superior visualization of peri-implant bone structures with fewer artifacts and improved diagnostic accuracy.