Related Experiment Video
Updated: Sep 11, 2025

Rejection of Fluorescence Background in Resonance and Spontaneous Raman Microspectroscopy
Published on: May 18, 2011
Raman measurements using an acousto-optic tunable filter spatial heterodyne Raman spectrometer with an echelle-mirror
None:
This study presents an acousto-optic tunable filter technology-based echelle-mirror spatial heterodyne Raman spectrometer (AOTF-EMSHRS) that integrates echelle gratings with mirror spatial heterodyne detection. The instrument uses the grating's high resolution and the AOTF's rapid wavelength selection capability to optimize the optical path through a mirror design. This results in a high signal-to-noise ratio (SNR), high spectral resolution, and broad wavelength range measurements. After calibration, the theoretical resolution reached 1.53 cm-1, with a single-order spectral detection range of 1574.3 cm-1. The spectral detection range spans from 100 cm-1 to 4397 cm-1. Experimentally, we detected and analyzed various organic and inorganic substances and solutions. We compared EMSHRS performances with and without the AOTF and found that the SNR of the AOTF-EMSHRS system has improved by at least twofold. Additionally, we measured solutions with different molar concentrations, various sulfate mixtures, and mixtures in transparent containers made from different materials, obtained comprehensive spectral information, and conducted detailed analyses. The AOTF-EMSHRS demonstrated excellent measurement performance during complex sample detection and analysis.
More Related Videos
09:46Direct Comparison of Hyperspectral Stimulated Raman Scattering and Coherent Anti-Stokes Raman Scattering Microscopy for Chemical Imaging
Published on: April 28, 2022
09:57Ultrafast Time-resolved Near-IR Stimulated Raman Measurements of Functional π-conjugate Systems
Published on: February 10, 2020
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...
IR Spectrometers