EARL compliance on the Biograph Vision Quadra PET-CT: phantom study for static and continuous bed motion acquisitions
Beverley F Holman1,2, Tamar Willson1, Bruno Ferreira1
1Nuclear Medicine, Royal Free Hospital, London, United Kingdom.
Frontiers in Nuclear Medicine
|August 13, 2025
Summary
Achieving EARL standards on advanced LAFOV PET scanners like the Biograph Vision Quadra reduces image quality, particularly in Ultra-High Sensitivity mode. A new "EARL 3" standard is proposed for modern PET systems.
Area of Science:
- Medical Imaging Physics
- Nuclear Medicine Technology
- Quantitative PET Imaging
Background:
- Long axial field-of-view (LAFOV) PET systems offer enhanced sensitivity.
- Compliance with European Association of Nuclear Medicine Research Laboratory (EARL) standards is crucial for quantitative PET imaging.
- The impact of EARL standards on advanced LAFOV PET systems, including the Siemens Biograph Vision Quadra, across various acquisition modes is not well-established.
Purpose of the Study:
- To identify reconstruction parameters for EARL 1 and 2 compliance on the Biograph Vision Quadra PET system.
- To evaluate compliance in both static and continuous bed motion (CBM) acquisitions across High Sensitivity (HS) and Ultra-High Sensitivity (UHS) modes.
- To optimize image quality while adhering to quantitative accuracy standards.
Main Methods:
- Utilized the IEC body phantom with 18F-FDG and varied acquisition modes (static, CBM, HS, UHS).
- Reconstructed images using standard and varied parameters (Gaussian filters, matrix sizes).
- Assessed EARL compliance using the EARL tool, focusing on SUV recovery coefficients (RCSUVmean, RCSUVmax, RCSUVpeak).
Main Results:
- Identified EARL-compliant reconstruction parameters for all tested acquisition modes.
- No significant differences in compliance were observed between static and CBM acquisitions.
- EARL compliance necessitated substantial image quality reduction, more pronounced in UHS mode and for EARL 1, resulting in smoother images with reduced contrast.
Conclusions:
- EARL compliance, while ensuring standardization, significantly compromises image quality on advanced LAFOV PET systems.
- The current EARL standards may not be suitable for the high-performance capabilities of modern PET scanners.
- An updated "EARL 3" standard is recommended to accommodate the advancements in LAFOV PET technology.